<7. Inspection and Maintenance>
7-6
IM 11M10B01-01E
(4) Insert the ring
1
and holding plate
4
in this order into the sensor. Hold the sensor with
fingers and rotate the joint
2
right and left to push it into the sensor slowly.
Lead wire connector
3
Heater
Sensor
1
4
2
(5) Rotate and tighten the ring
1
into the joint
2
. Secure the holding plate
4
with two
screws (7).
Connect the lead wire connector of the sensor to the PCB board.
Lead wire connector
Heater
Sensor
7
Sensor constant tag
3
1
4
2
(6) When the installation of the sensor is finished, take a note of the sensor constant on the
sensor constant tag attached to the sensor lead wire, and then close the cover.
(7) Turn on the power, and when “HEAt” appears and countdown starts, enter sensor constant
“SEnS.” As for how to enter it, see 6.1.7 “Setting Sensor Constant.”
Be sure to perform calibration after replacing the sensor.