<7. Inspection and Maintenance>
7-3
IM 11M10B01-01E
7.3
How to Replace Sensor
When the sensor has deteriorated, replace it with a new sensor. As for the part number, see
CMPL at the end of this manual. Be sure to perform calibration after replacing the sensor.
Replace the sensor according to the following instructions.
7.3.1
Removing Sensor
(1) Turn off the power, and be sure to pull the power cord out of the socket.
(2) Wait for about 1.5 hours until the heater temperature falls to the ambient temperature level.
WARNING
Be sure to replace the sensor after the heater temperature has fallen sufficiently. Not doing so
may cause burn injuries.
(3) To remove the upper cover, remove set screws at three points on the rear panel of the
OX400.
(4) Slide the upper cover backwards and remove it upwards.
WARNING
AVERTISSEMENT
Rear panel
Fan
Sensor
Heater
Side panel
When the upper cover has been removed, you can see the heater sensor.
Remove the three screws and slide the upper cover backwards.
!
INLET
OUTLET
N200
C
US
R
172608
SERIAL
(RS-232)
FUSE
250V T3.15A
!
RCCOM
RC3
RC2
RC1
FAIL
DO
PUMP
OFF
G
V
-
+
mA
-
+
MS1
MS2
MS3
MS
COM
MSI
COM
MSI2
MSI1
Figure 7.1
Removing Cover
(5) Remove the lead wire connector of the sensor from the PCB board.
(6) Loosen the joint ring
1
on the sensor outlet side and remove it towards the heater side.
(7) Pull out the joint
2
while holding the sensor.