<2. Installation >
2-10
IM 11M10A01-01E
2.7
Calibration Kit (Optional)
The calibration kit for the OX100 oxygen analyzer comprises a standard gas cylinder and
pressure regulating valve to supply standard gas for calibration, when necessary.
Standard
Gas
OX100 Sensor
K9424JD
(255)
(310)
(90)
∅
75
Pressure
Gauge
Gas Vent Tubing
Needle Valve Assembly
Needle Valve Assembly, with Pressure Reducing Valve
and Fixed Throttle
Rc1/8 (Nominal diameter of piping
connection for K9424JD)
Standard Gas Cylinder
Standard Calibration
Gas (6 bottles/box)
K9424JC
Approx. 340
Approx. 180
Part Number
Specifications
K9424GA
Sampling unit
K9424JA
Calibration kit: 1000 ppm, 6 cylinders (1L) + needle valve assembly, with flow adjustment
K9424JB
Calibration kit: 1000 ppm, 6 cylinders (1L) + needle valve assembly, without flow adjustment
K9424JR
Calibration kit: 10000 ppm, 6 cylinders (1L) + needle valve assembly, with flow adjustment
K9424JS
Calibration kit: 10000 ppm, 6 cylinders (1L) + needle valve assembly, without flow adjustment
Figure 2.10
Calibration Kit
CAUTION
Shelf life of gas cylinder
The shelf life of a calibration gas shall be one year from the manufacturing date specified on the
label of the cylinder. Yokogawa shall not guarantee accuracy of the equipment, if it is calibrated
with an expired calibration gas.
NOTE
Calibration gases are not available from Yokogawa. Consult with local suppliers.
4th Edition : Oct. 23, 2019-00