background image

  

<1.  Overview>

1-1

IM 11M10A01-01E

1.  Overview

1.1 

Cautions Regarding Use

CAUTION

·  This is a precision instrument; be careful not to drop it.
·  Do not disassemble it, as it may stop operating normally.
·  As far as possible, mount horizontally. Secure the sensor so that it can’t vibrate. Avoid 

installing in places where the temperature exceeds 70°C.

·  Do not mount this instrument in a place that is exposed to direct sunlight, or to direct 

radiation from a furnace, boiler or the like, in a windy place, or in a location that is subject to 

vibration.

·  Avoid mounting in a location where there may be corrosive gases, gases other than the 

measured gas, or organic silicon vapor.

·  Where the furnace gas of a nitrogen reflow furnace contains organic solvent at a few ppm to 

percentage level, the sensor should be used with the optional sampling system (K9424GA), 

to avoid contact with the solvent.

·  This instrument was designed to be resistant to electrical noise, but to avoid invalid readings 

it would be wise to avoid using electrically noisy devices in the vicinity.

·  The sensor is a consumable. Its life depends on usage and environment, but is about one 

year.

·  Be sure to keep this manual in a handy place when using this instrument.

1.2 

Standard Specifications

Measured gas:   

 Measures O

2

 in N

2

 or mixed gases; can’t contain silicone vapor, 

flammable gas or organic solvent.

Measuring principle:   Current-limited Zirconia sensor
Measuring range:  

0 to 1000 ppm O

2

, 0.1 to 25% O

2

 (measuring range autoswitching) 

 

 

Note:  If you want an oxygen (concentration) control range of 0.2 to 0.5%,please consult with us, or your Yokogawa 

salesperson or representative.

Sampling method:  

Natural diffusion (direct input)

Linearity : 

 

 ±5% of span (for 0-1000 ppm range, with O

2

 being in the range 0 to 100 

ppm). ±3% of span (for 0-1000 ppm range, with O

2

 being in the range 

100 to 1000 ppm). ±2% of span (for 0.1 to 25% range, with O

2

 being in 

the range 100 to 1000 ppm).

Repeatability :    

 ±5% of span (for 0-1000 ppm range, withO

2

 being in the range 0 to 100 

ppm). ±2% of span (for 0-1000 ppm range, with O

2

 being in the range 

100 to 1000 ppm). ±2% of span (for 0.1 to 25% range, with O

2

 being in 

the range 100 to 1000 ppm). 

Stability (Drift):   

±2% of span per week.

Display: 

 

LED digital display

Output signal:   

4-20 mA DC isolated (measuring range: 0 to 1000 ppm or 0.1 to 25%).

Contact outputs (relay) :  Closed if fail, contact rating 250 V AC / 3A 

Closed if alarm, contact rating 250 V AC / 3A.

4th Edition : Oct. 23, 2019-00

Summary of Contents for OX100

Page 1: ...User s Manual OX100 General Purpose Oxygen Analyzer IM 11M10A01 01E IM 11M10A01 01E 4th Edition...

Page 2: ...he instrument hardware or software from being damaged or a system failure from occurring The following are signal words to be found only in our instruction manuals CAUTION This symbol gives informatio...

Page 3: ...rnally It is recommended that you ask the local service representative to change the fuse do not either change it yourself or open the case to check the fuse Turn off the power switch and disconnect t...

Page 4: ...wa does not authorize Failure due to power supply voltage frequency being outside specifications or abnormal Failure caused by any usage out of scope of recommended usage Any damage from fire earthqua...

Page 5: ...maintenance which is determined by gas conditions Filters and activated carbon packs should be replaced at intervals which depend on the operating conditions For best measurement use new activated car...

Page 6: ...2 4 Wiring 2 6 2 4 1 Power Wiring 2 6 2 4 2 Connecting to an External Instrument 2 7 2 5 Installation of Sensor 2 7 2 6 Installing the Optional Sampling Unit K9424GA 2 8 2 7 Calibration Kit Optional...

Page 7: ...he Needle Valve Assembly to the Gas Cylinder App 5 2 Supplying the Calibration Gas App 6 3 Stopping Supplying the Calibration Gas App 6 Customer Maintenance Parts List CMPL 11M10A01 01E Revision Infor...

Page 8: ...environment but is about one year Be sure to keep this manual in a handy place when using this instrument 1 2 Standard Specifications Measured gas Measures O2 in N2 or mixed gases can t contain silic...

Page 9: ...ure 0 to 70 C gas flow up to 1 m s no corrosive gases like SO2 can t be used in atmosphere with flammable gases corrosive gases or silicone vapor Power consumption up to 15 VA Power supply 100 V AC 10...

Page 10: ...424JA Calibration kit Six 1 liter flasks 1000 ppm O2 N2 gas outlet cock with flow adjuster K9424JB Calibration kit Six 1 liter flasks 1000 ppm O2 N2 gas outlet cock without flow adjuster K9424JR Calib...

Page 11: ...ent Body 96 26 2 23 13 100 91 6 96 91 6 Body only Part no K9424A 1 to 10 Bracket Part no B9877BA Panel thickness Unit mm Fig 1 1 External Dimensions of Instrument Panel Dimensions 92 0 8 0 92 0 8 0 25...

Page 12: ...gure 1 2 Sensor dimensions Sample outlet Rc1 4 Sample inlet Rc1 4 220 136 146 290 240 Power line connector Needle valve Activated carbon filters Flowmeter Power line switch Connector to joint sensor S...

Page 13: ...eedle valve assembly with flow adjustment K9424JS Calibration kit 10000 ppm 6 cylinders 1L needle valve assembly without flow adjustment Figure 1 4 Calibration Kit Optional 1 5 Sensor Operating Princi...

Page 14: ...in 1000 ppm O2 calibration mode 4 20 mA lamp Lit for range 0 1000 ppm O2 and 4 20 mA output PPM lamp Lit when range is 0 to 1000 ppm O2 lamp Lit when range is 0 1 to 25 O2 FAIL lamp Self diagnostic f...

Page 15: ...able color Red Sensor White Sensor Yellow Heater Black Heater Figure 2 3 Terminal layout Terminal layout ALM AIR SET FAIL 4 20mA ppm 1000ppm ENT MODE Enter key Enters current mode setting Mode key Inc...

Page 16: ...ons Switch to Cal1 mode Increase setting Decrease setting Enters setting and switches to measure Switch to Cal2 mode NA NA Switch to measure Starts calibration At end switches to measure Increase sett...

Page 17: ...e affected if the installation location is subject to vibration 4 Horizontal location As far as possible the installation location should be horizontal Avoid locations with the following conditions CA...

Page 18: ...in Fig 2 5 mount this instrument on the panel using the supplied brackets The brackets mount on the top and bottom of the case at the specified positions see Fig 1 1 External Dimensions of Instrument...

Page 19: ...g power cable To minimize the possibility of fire use 600 V rated heavy vinyl insulated cable JIS C 3307 or similar for power wiring Before applying power connect the ground terminal to a ground conne...

Page 20: ...nals 13 14 relay contact closes if sensor heater element goes open circuit WARNING To minimize the possibility of electrical shock ensure that power source is off before wiring power cable 2 5 Install...

Page 21: ...e space for inspection and maintenance such as replacing filter element is secured The unit should be securely mounted on a stand The sampling unit incorporates a suction pump which requires power sup...

Page 22: ...pply 100 V 50 60 Hz Piping connection Rc1 4 Flowmeter Sensor Connector for 8 mm O D Needle Valve OX100 Sensor Activated Carbon Filter Switch 3 Sample Inlet Rc1 4 Rc 1 4 Sample Gas Inlet Sample Gas Out...

Page 23: ...Sampling unit K9424JA Calibration kit 1000 ppm 6 cylinders 1L needle valve assembly with flow adjustment K9424JB Calibration kit 1000 ppm 6 cylinders 1L needle valve assembly without flow adjustment K...

Page 24: ...ing order first perform one point calibration with air then 1000 ppm O2 calibration Before calibrating turn on the power and allow the OX100 to warm up for at least 10 minutes then perfom air cleaning...

Page 25: ...arm lamp lights for 2 sec when you attempt calibration replace the sensor part no K9424BA CAUTION Ideally you should use 1000 100 ppm O2 N2 standard gas for calibration Gas flow should be between 800...

Page 26: ...eases Alarm Setting 3 3 Alarms and Alarm Output If the O2 concentration reaches the High Low Alarm setting entered in Section 3 2 above the ALM lamp lights and the alarm output contact between termina...

Page 27: ...Blank Page...

Page 28: ...n the specified pressure range 2 Clean probe tip Large errors 1 Temperature out of range 2 High pressure or negative pressure around sensor 3 Oxidation reduction or high concentration of corrosive gas...

Page 29: ...Blank Page 4th Edition Oct 23 2019 00...

Page 30: ...o the factory For repair or maintenance contact your Yokogawa representative 1 Specifications Pressure at sampling point 1013 40 hPa Pressure at sample outlet 1013 40 hPa Gas temperature 50 C max Orga...

Page 31: ...sampling tubing from the sampling point through a stop valve to the sample gas inlet GAS IN with care not to produce gas leakage The piping connection of the sample gas inlet is Rc1 4 Use an appropria...

Page 32: ...egulating valve of the flowmeter and the needle valve on the front completely by turning the valves counterclockwise completely 5 Turn on the power switch of the unit The suction pump starts working 6...

Page 33: ...ack into the case until the top of the pack sinks under the brim Thrust the pipe protruding from the back of the lid onto the top center not on the seam and tighten the lid securely Figure 3 5 Connect...

Page 34: ...are either one according to your application 1 Connecting the Needle Valve Assembly to the Gas Cylinder Take the following steps to connect the gas cylinder to the needle valve assembly 1 Turn the han...

Page 35: ...ntil it can no longer turn so that the needle tip pokes into the gas cylinder See Figure 3 4 Turn the handle of the needle valve counterclockwise to the position shown in Figure 1 5 Turn the pressure...

Page 36: ...se Oxygen Analyzer CMPL11M10A01 01E 1st Edition Dec 2000 YK 4th Edition Feb 2019 YK 1 2 Item Part No Qty Description 1 K9424BA 1 Sensor 2 K9424CA 1 Cable All Rights Reserved Copyright 2000 Yokogawa El...

Page 37: ...7 Item Part No Qty Description 1 L9866CX 1 Flow Meter 2 K9424GN 1 Pump Assy 3 K9424HE 1 Power Unit 4 K9424GZ 1 Filter Assy 1 5 K9424GY 1 Filter Element active carbon filter 5 packs 6 E7050AC 1 Valve 7...

Page 38: ...ion Feb 2019 YK All Rights Reserved Copyright 2000 Yokogawa Electric Corporation CMPL 11M10A01 01E K9424JC NEEDLE VALVE ASSY 1 2 Item Part No Qty Description 1 K9424JD 1 Needle Valve Assy 2 K9424HQ 1...

Page 39: ......

Page 40: ...arding Use Added an item p 1 2 Section 1 2 Standard Specifications Changed operating temperature range p 1 3 Added Section 1 3 MS Code p 1 5 Section 1 3 External Dimensions Figure 1 2 Changed p 1 6 Se...

Page 41: ...Blank Page...

Reviews: