2.  INSTALLATION AND ENVIRONMENTAL 

CONDITIONS

 • Avoid the following environments for installation locations:  

Areas with vibrations, corrosive gases, dust, water, oil, 

solvents, direct sunlight, radiation, a strong electric field, 

and/or a strong magnetic field, direct radiant heat, wind, 

temperature fluctuation.

 • If there is any risk of a surge being induced into the power 

line and/or signal lines due to lightning or other factors, a 

dedicated lightning arrester should be used as protection for 

both this converter and a field-installed device.

Environmental Conditions

Operating 

temperature range

0 to 50°C

Operating humidity 

range:

5 to 90% RH (no condensation)

Ambient Condition Avoid installation in such environments as 

corrosive gas like sulfide hydrogen, dust, 

sea breeze and direct sunlight. 

Use indoors at an altitude of 2000 m or less.

Power Supply and Isolation

Power Supply 

Rated Voltage: 

24 V DC  or 100-240 V AC/DC   50/60 Hz

Power Supply 

Input Voltage:

24 V DC   (±10%) or 100-240 V AC/DC    

(−15%, +10%) 50/60 Hz

Power 

consumption:

1.8 W at 24 V DC; 1.5 W at 110 V DC

2.6 VA at 100 V AC, 4.0 VA at 200 V AC

Insulation 

resistance:

100 MΩ minimum at 500 V DC between 

Ethernet, RS-485, power supply and 

grounding terminals mutually

Withstanding 

voltage:

1000 V AC for one minute between Ethernet 

and RS-485 terminals mutually 

2000 V AC for one minute between 

(Ethernet, RS-485), power supply, and 

grounding terminals mutually

3.  EXTERNAL WIRING

 WARNING

Be sure to turn OFF the power supply before wiring to 

avoid the risk of electric shock. Use a tester or similar 

device to ensure that no power is being supplied to a 

cable to be connected.

Wiring should be connected to the terminals on the socket of the 

product. The terminals for external connections are of M3 screws. Use 

crimp-on terminal lugs for connections to the terminals.

 • Recommended cables: A nominal cross-sectional area of 

0.5 mm

2

 or thicker for signal cables, and that of 1.25 mm

2

 or 

thicker for power cables.

Wiring Diagram

    

10

11

2

5

6

8

Power supply

11

10

8

L+

N−

GND

B+

A−

COM

5

2

6

RS-485

communication

Terminator

2

5

6

2

5

6

2

5

6

Terminating

resistor

220 Ω

Terminating

resistor

220 Ω

Grounding

B+

VJET

VJ series

Signal Conditioner

VJ series

Signal Conditioner

A−

COM

(SG)

B+

A−

COM

(SG)

B+

A−

COM

(SG)

  CAUTION

 

Adhere strictly to the specifications to avoid 

overheating or damage. Before turning on the 

power, ensure the following:

 

Power supply voltage and input signal value 

applied to the product should meet the required 

specifications.

 

The external wiring to the terminals and wiring to 

ground are as specifications.

 

Do not operate the product in the presence of 

flammable or explosive gases or vapors.

 

This product is sensitive to static electricity; 

exercise care in handling. Before you operate the 

product, touch a nearby metal part to discharge 

static electricity.

 

The power line and input/output signal lines should 

be installed away from noise-generating sources. 

Otherwise performance cannot be guaranteed.

 

Make sure to earth ground the ground terminal 

through minimum resistance. The length and 

thickness of the grounding cable should be as short 

and thick as possible. Directly connect the lead from 

the ground terminal (terminal no. 8) of the product 

to the ground. Do not carry out daisychained inter-

ground terminal wiring

 • Connet the VJET to the host device using either 10BASE-T 

or 100BASE-TX.  The VJET detects 10BASE-T or 100BASE-

TX automatically.  10BASE-T/100BASE-TX are 10 Mbps/100 

Mbps Ethernet standard using twisted-pair cables.  In 

10BASE-T/100BASE-TX networks, personal computers and 

other host devices are connected in a star pattern through a 

hub.

Link LED

Connector for 

10BASE-T/100BASE-TX

Ethernet

HUB

or

Devices that can be connected to Ethernet

PC

Touch panel

Activity LED

Terminator

Link LED (Lower side)

Color

Off

Amber

Green

Meaning

No Link

10 Mbps

100 Mbps

Activity LED (Upper side)

Color

Off

Amber

Green

Meaning

No Activity

Half-Duplex

Full-Duplex

The VJET has Link LED (lower side) and Activity LED (upper 

side) on the connector part of the front. These LEDs turn on 

in green or amber. (see the External Dimensions.)