<6. TROUBLE SHOOTING>
6-1
IM 11G02Q02-01EN 2nd Edition : Sep. 04, 2020-00
6. TROUBLE SHOOTING
If the communication is unavailable, check the following items.
• Whether all devices related to communication are turned on.
• Whether connections are correct.
• Whether the number of connected instruments and connection distance are as specified
• Whether communication conditions coincide between the master station (host computer)
and slave stations (instrument)
Transmission speed : 38400bps
Data length : 8 bits
Stop bit : 1 bit
Parity : None
Whether send/receive signal timing conforms to Section 5.6 in this manual.
• Whether the station No. designated as send destination by the master station coincides with
the station No. of the connected instrument.
• Whether more than one instrument connected on the same transmission line shares the
same station No.
• Whether the station No. of instruments is set at other than 0.
If it’s 0, the communication function does not work.