background image

IM 12B07K02-01E

 

 

< 1. Specification >

1-1

1.  Specification

The Model FU20 offers a simple and cost effective solution for a wide variety of waste water and 

process applications. This all-in-one sensor provides simultaneous measurement of pH, ORP 

and temperature. The rugged Ryton body is designed for easy installation into on-line via the 3/4 

inch tapered threaded connections provided on both ends of the sensor.

1.1 

Standard Specifications

Measuring range: 

pH; 2 to 12   ORP; -1500 to 1500 mV

Solution temperature :  -5 to 105°C (sensor only, sample is not freezed)
Solution pressure :  

0 to 1 MPa (sensor only)

Solution conductivity :  50 µS/cm or higher
Solution flow velocity :  2 m/s max.
Internal electrolyte: 

Saturated KCl (Non refillable)

Temperature sensor :  Pt1000 (Note)

Body materials:  

Ryton (PPS 40GF)

Weight:  

 

Approx. 0.5 kg (Body)

Storage temperature:  -10 to 50 °C

Applicable Liquid Analyzer/converter:   FLXA202/FLXA21 2-Wire Liquid Analyze,  

FLXA402 4-Wire Converter, PH450G pH/ORP converter

Applicable holder:  

 Special order for PH8HG, HH350G with optional adapter, cannot be 

used with HH350G/-JT.

Note:  The temperature sensor included in the FU20 is intended to provide indication and cell compensation. The construction has not 

been tested to the pressure vessel standards required for plant temperature control.

NOTE

Cautions for using the FU20 sensor

Note that the FU20 sensor has limited life which depending on installation process.
Be careful not to allow liquid junction and glass of the pH sensor to dry out.
When storing the sensor, be sure to attach the protective cap (supplied with it on shipment) to its tip.
The sensing glass membrane of the pH sensor gradually deteriorates during storage.
A virgin pH sensor may become unusable because of its deterioration. Avoid storing the sensor 

for a long time (Use within a year is recommended).

1.2 

Model and Suffix codes

Model

Suffix Code

Option Code

Description

FU20

................................... ...................

Wide body Sensor

Type

-05

-10

...................

...................

Fixed cable 5 meter

Fixed cable 10 meter

Temperature element

-T1

................... Pt1000

Model

-NPT

................... Dome shape model

Note:  Special order is needed for using with PH8HG drop-in guide pipe.

4th Edition : Nov.20,2018-00

Summary of Contents for FU20

Page 1: ...User s Manual Model FU20 pH ORP Sensor IM 12B07K02 01E IM 12B07K02 01E 4th Edition...

Page 2: ...g outer box and then immediately notify the carrier and the relevant Yokogawa sales office Make sure the Model Code on the sensor is the same as on the packing list For details of the Model Code refer...

Page 3: ...fore using the product The purpose of these user s manuals is not to warrant that the product is well suited to any particular purpose but rather to describe the functional details of the product No p...

Page 4: ...to which the sensor have been exposed n In the following cases customer will be charged repair fee regardless of warranty period Failure of components which are out of scope of warranty stated in inst...

Page 5: ...Blank Page...

Page 6: ...r the safe use of this equipment ii After sales Warranty iii 1 Specification 1 1 1 1 Standard Specifications 1 1 1 2 Model and Suffix codes 1 1 1 3 External Dimensions 1 2 2 Installation and Wiring 2...

Page 7: ...Blank Page...

Page 8: ...verter Applicable holder Special order for PH8HG HH350G with optional adapter cannot be used with HH350G JT Note The temperature sensor included in the FU20 is intended to provide indication and cell...

Page 9: ...13 14 11 12 150 19 6 22 5 6 5 19 6 25 9 PG7 Cable length 5 m or 10 m Standard type FU20 T1 NPT Sensor body PPS resin Note Can be fixed to 3 4NPT female screw 3 4 inch tapered thread Note 3 4 inch tap...

Page 10: ...an immersion assembly For best results the FU20 should be mounted with the process coming towards the sensor and positioned at least 15 above the horizontal plane to eliminate air bubbles in the glass...

Page 11: ...REFERENCE SOLUTION GROUND GLASS Measure GE SHIELD 11 12 13 14 15 16 Wire Marking Figure 2 3 Sensor Wiring for pH measurement For ORP measurement PG7 FLXA202 FLXA21 FLXA402 PH450G 2 REFERENCE SOLUTION...

Page 12: ...tion clean off remaining stains by rinsing with water Stains due to Oily Materials Wash off stains by submerging in a neutral detergent solution in a beaker etc for from several tens of minutes to sev...

Page 13: ...Blank Page...

Page 14: ...code of NAG deleted for Model and suffix codes FU20 P 1 2 Some revision of external dimensions P 2 1 Some revision of Installation and In line Installation added screw in torque changed P 2 2 Some re...

Page 15: ...Blank Page...

Reviews: