IM 12B07K02-01E

  

< 3. Maintenance >

3-1

3. Maintenance

The sensor has limited life. If buffer solution calibration becomes impossible due to deterioration 

of the sensor, replace it with a new one. After the sensor is replaced, always carry out a buffer 

solution calibration.

NOTE

A glass electrode whose glass membrane has dried out does not provide stable readings until 

it gets used to the solution. Place in advance the glass electrode in the 3 mol/L KCl solution for 

several hours. When storing the sensor, use a sensor cap (supplied with it on shipment) to its tip.

n

  Cleaning Electrode and Liquid Junction

Staining of an electrode or liquid junction can cause measurement errors. Therefore, if he 

measured solutions tend to stain the electrode, the electrode and liquid junction must be cleaned 

periodically - depending on the degree of staining.
To clean the sensor, proceed as follows:
(1)  Take out the sensor.

(2)  Clean the electrode or liquid junction as follows. Different cleaning solutions may be 

required depending on the staining.

●  Stains due to suspended Solids, Sticky Materials, Microbes or the like
  Using soft tissue paper, wipe the stains off the electrode or liquid junction. In addition, clean 

off remaining stains by rinsing with water.

●  Stains due to Oily Materials
  Wash off stains by submerging in a neutral detergent solution in a beaker, etc, (for from 

several tens of minutes to several hours depending on the degree of staining).

●  Chemical Stains such as due to Metallic Adsorption

  Place the electrode or liquid junction in a diluted hydrochloric acid solution (1 to 2%) for 

several minutes (acid washing). Afterwards rinse the sensor with water.

●  Protein Deposits

  It should be removed with a protein enzymatic solution. For instance a solution containing 

8.5 mL concentrated hydrochloric acid and 10 gram pepsin in 1 liter water will do.

(3)  Check deterioration of the sensor.
(4)  Calibrate the sensor with buffer solution.

(5)  Mount the sensor again.

CAUTION

Avoid using non-polar solvent like tri-chloro ethylene, toluene or hexane. Cleaning with iso-

propanol or methanol is acceptable.
The non-polar solvents will break up the gel-layer on the glass bulb and requires that the 

electrode to be soaked in water for at least 12 hours before functioning normally again. 

The Teflon (PTFE) diaphragm of the sensor can sometimes be regenerated by putting it in hot 

(approx. 70°C) 3 molar Potassium Chlorine (KCl) solution and letting it cool to room temperature.

For details of calibration procedure, refer to relevant user’s manual of transmitter/converter.

4th Edition : Nov.20,2018-00

Summary of Contents for FU20

Page 1: ...User s Manual Model FU20 pH ORP Sensor IM 12B07K02 01E IM 12B07K02 01E 4th Edition...

Page 2: ...g outer box and then immediately notify the carrier and the relevant Yokogawa sales office Make sure the Model Code on the sensor is the same as on the packing list For details of the Model Code refer...

Page 3: ...fore using the product The purpose of these user s manuals is not to warrant that the product is well suited to any particular purpose but rather to describe the functional details of the product No p...

Page 4: ...to which the sensor have been exposed n In the following cases customer will be charged repair fee regardless of warranty period Failure of components which are out of scope of warranty stated in inst...

Page 5: ...Blank Page...

Page 6: ...r the safe use of this equipment ii After sales Warranty iii 1 Specification 1 1 1 1 Standard Specifications 1 1 1 2 Model and Suffix codes 1 1 1 3 External Dimensions 1 2 2 Installation and Wiring 2...

Page 7: ...Blank Page...

Page 8: ...verter Applicable holder Special order for PH8HG HH350G with optional adapter cannot be used with HH350G JT Note The temperature sensor included in the FU20 is intended to provide indication and cell...

Page 9: ...13 14 11 12 150 19 6 22 5 6 5 19 6 25 9 PG7 Cable length 5 m or 10 m Standard type FU20 T1 NPT Sensor body PPS resin Note Can be fixed to 3 4NPT female screw 3 4 inch tapered thread Note 3 4 inch tap...

Page 10: ...an immersion assembly For best results the FU20 should be mounted with the process coming towards the sensor and positioned at least 15 above the horizontal plane to eliminate air bubbles in the glass...

Page 11: ...REFERENCE SOLUTION GROUND GLASS Measure GE SHIELD 11 12 13 14 15 16 Wire Marking Figure 2 3 Sensor Wiring for pH measurement For ORP measurement PG7 FLXA202 FLXA21 FLXA402 PH450G 2 REFERENCE SOLUTION...

Page 12: ...tion clean off remaining stains by rinsing with water Stains due to Oily Materials Wash off stains by submerging in a neutral detergent solution in a beaker etc for from several tens of minutes to sev...

Page 13: ...Blank Page...

Page 14: ...code of NAG deleted for Model and suffix codes FU20 P 1 2 Some revision of external dimensions P 2 1 Some revision of Installation and In line Installation added screw in torque changed P 2 2 Some re...

Page 15: ...Blank Page...

Reviews: