<1. Operation>
1-1
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
1. Operation
After every wiring is completed, power on the converter. LCD screen first displays startup screen
, then it transitions automatically “Home screen”.
When single measurement is set to display on a screen, Home screen will be skipped and Main
screen will show up. See chapter 2 to know screens.
or
Home Screen
Main Screen
Figure 1.1
Sample image of startup screens
NOTE
It takes more than ten seconds for the startup screen to switch to Home screen after FLXA402 is
powered on.
When you use SENCOM SA and connect no SENCOM SA or incompatible sensors, an error
appears to indicate inappropriate setting or measurement error. To review and modify initial
setup, if you need. See chapter 4.
NOTE
For several seconds after power is on, 4 mA is output. This does not imply the device has fault.
NOTE
In the beginning of the measurement, you might see the measurement displayed “----”. This
happens because the sensor is preparing for the measurement and not because of fault. Wait
until the sensor is ready.
While the measurement is displayed “----”, mA output follows the order defined in mA output
settings: mA output settings > mA > mA (Configure Hold) > Last or Fixed
NOTE
When there is no touch operation for 10 minutes, the screen automatically switches to Monitor
screen