<4. Setting converter>
4-24
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
4.5.4
Others
Configure other settings of converter display.
l
Auto Return
The converter reverts to the Monitor screen (if Monitor screen is set to “Off”, reverts to Main
screen), when there is no touch operation during the set time interval.
You can select the interval to implement Auto Return among “Disable”, “10 min”, “60 min”. If you
want to disable this function, select “Disable”.
NOTE
The auto return does not work on the following screen: Trend, Calibration, sensor configuration.
NOTE
The default of Auto return is 10 min. If you expect to have a time-consuming operation such as
calibration, “60 min” or “Disable” is recommended. However, if you do not save the data after
changing the setting, auto return is not activated.
l
Backlight saver
To keep LED life longer, you can set backlight saver to turn off backlight automatically. When
there is no touch operation for set interval, the backlight will be off. Select Backlight saver
“Disable”, “10 min”, “60 min.”. When Backlight is off, once you touch the screen, the back light will
be turned on.
When Backlight is off but a fail happens, the backlight starts flashing and keeps flashing until the
fail is cancelled.
While Backlight flash (different function from the one mentioned above, see the next page) is
being activated, backlight will flash and never go off automatically.
l
MONITOR Display
Set automatic screen shifting to Monitor screen. Select “Off” “On”. Factory default is “On”. Monitor
screen does not come out when a fail happens, Hold is on, or Wash is in progress.
When this setting is “On”, the screen auto-returns to Monitor screen.
l
Alarm NE107
When a fail on the converter or sensors is detected, Home or Main screen displays the diagnostic
information.
Set “On” or “Off” to Alarm NE107. The default is “On”.
Main alarm
, , ,
or
,
Sensor alarm
, ,
, or
When “Alarm NE107”* is “On”, (“On” as factory default), the icons in accordance with NAMUR
NE107*, one of
, , ,
lights on upon detection of each alarm.
If a sensor alarm was detected, one of
, ,
, lights on.
When “Alarm NE107” is “Off”, only
(Fault) or
(Warning) blinks.