<4. Setting converter>
4-9
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
4.3
mA output settings
Configure the settings of mA output.
n
mA1/mA2/mA3 (Ad)/mA4 (Ad)
Set each mA output. “Output” is default.
Output:
mA output according to each setting
Simulate
output fixed mA set by output span (%)
Off
No mA is output. 4 mA is fixed output.
When mA Output/Input is “-A2”, mA output of only mA1 and mA2 are settable. When mA Output/
Input is “-A4”, mA output of mA1, mA2, mA3(Ad), mA4(Ad) are settable. When mA Output/Input
is “-A2”, setting mA3(Ad)/mA4(Ad) does not affect the operation.
NOTE
Set “Output” at mA1 for HART.
4.3.1
Output
l
Process parameter slot
FLXA402 allows multiple sensor connection, according to specification you choose.
Select a sensor from “Sensor 1-1”, “Sensor 1-2”, “Sensor 1-3”, “Sensor 1-4”, “Sensor 2-1”,
“Sensor 2-2, “Sensor 2-3”, “Sensor 2-4”, “Converter” to output process parameter. If you
set converter parameter (ex. mA input, script, arithmetic by multiple sensors) to the process
parameter, select “Converter.”
NOTE
Latest revision does not support script, or arithmetic by multiple sensors.
Sensors must be assigned addresses for MODBUS network in advance. Read 4.6.1.