Table 6.4 DEVICE_STATUS_2 (System Fault)
Error
Description and Remedy
NAMUR
NE107
category
DIAGNOSIS
(Classic)
DIAGNOSIS
(Condensed)
DIAGNOSIS
EXTENSION
Wash response time
failure
Not used.
—
—
—
—
Calibration Due
System was not maintained
within the preset period.
Perform maintenance. Increase
Calibr. Interval.
M
Maintenance
required
Maintenance
required
Calibration
Due
Outputs in HOLD
Press HOLD in main display.
Contact your local sales office in
case the HOLD flag reappears.
—
—
—
—
mA output burn high Upscale burnout situation.
Indication of sensor fault.
Check measured process values.
—
—
—
—
mA output burn low Downscale burnout situation.
Indication of sensor fault.
Check measured process values.
—
—
—
—
Error in mA table
mA table is not properly defined.
Execute 'Check values' on the
display of FLXA21.
—
—
—
—
mA calculation error Problem with calculating a mA
value.
Check the 'Process parameter'
for mA.
—
—
—
—
mA configuration
error
Problem with mA and Process
parameter.
Select a correct 'Process
parameter' for mA on the display
of FLXA21.
F
Configuration
not valid
Failure of
the device or
armature
Configuration
Error
Internal com. Error
Contact your local sales office.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure
Chksum err.(CPU
AS)
Contact your local sales office.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure
EEPROM err.(CPU
AS)
Write or read error in the
EEPROM.
Contact your local sales office.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure