<1> Wiring
Make connection as shown in the following figure.
<Note>
• When the 2P Test Leads are not used, short P and C
terminals.
#
WARNING
• Please be sure to use a voltage detector to check a common
earth of commercial power supply.
• Do not use the instrument to check a common earth of
commercial power supply.
A danger will be caused because the voltage may not be
displayed even in case of a live conductor when the connection
of the earth electrode to be measured has come off, or when the
connection of the test leads of the instrument is not correct etc.
• Do not use the instrument to measure the voltage of commercial
power supply.
The instrument is not designed for voltage measurement of
commercial power supply.
When using the 2P Test Leads, P and C terminals will be short-
circuited and the input impedance will be reduced. The residual
current circuit breaker may operate when making measurement
of the voltage in the circuit with the breaker.