10-24
IM DR231-01E
10.10 Selecting Display Update Interval, Registering
SET UP Menu, and Terminating SET UP Menu
D I S P
M O D E
P R I N T
F U N C
I N S
D E L
R A N G E A L A R M
S E T
C H A R T
E N T E R
F E E D
M. F U N C 1
M. F U N C 2
RECORD
ALARM
CHART
KEYLOCK
R E C O R D
SETUP=DISPLAY
↓
COLOR LOCK FUNC_PARM SET_PARM DISPLAY
↑
>Select Setting Parameter
E S C
• Enter the SET UP menu in the following procedure:
1. When the display presents the operation display mode or a menu other than the SET UP
menu, first turn off the power switch.
2. Turn the switch back on while pressing the DISP key and continue pressing the DISP key
for about 5 seconds.
• Select/set
using the
keys.
• To escape from a lower menu, press the MODE (ESC) key. Its main menu will appear,
although new settings/selections will not be kept.
• New settings/selections will be kept when you reach the final display of that setting.
AUTO INTERVAL=2
>2 3 4 5
***AUTO INTVL SET***
SETUP=ABORT
ABORT STORE
DISPLAY
END
ENTER
ESC
ENTER
ENTER
ESC
ENTER
( S E T U P e n d )
( M a i n m e n u )
( L o w e r m e n u )
SET UP=
E x p l a n a t i o n
Selection of Display Update Period (DISPLAY)
Select the display update interval when the operation display mode in the display is AUTO.
• Select 2, 3, 4, or 5 seconds.
• The default setting is 2 seconds.
Registration/Abortion of Setting/Selection Details and Termination of SET UP
menu (END)
In order to make the details set/selected in the SET UP menu effective, a registering operation is
necessary. In contrast, the status can also be brought to that before setting/selection (abort)
without registration. Terminate the SET UP menu after registering or aborting operation.
• ABORT: Registration is aborted and the SET UP menu is terminated.
• STORE: After registration, the SET UP menu is terminated.
• The default setting is ABORT.
Select the END menu, then press the ENTER key.
The operation display mode is reached about 10 seconds after pressing the ENTER key.
Summary of Contents for DR231
Page 12: ...10 IM DR231 01E...