IM CL250

the buzzer beeps when the reading is about 50Ω or less. Press the button 

again to exit the continuity check mode.

(7) 

 Button : Used for zero adjustment on 400A DC range or for reset-

ting the reading in the MAX mode. "

" symbol is shown on the display 

when zero adjustment is enabled on 400A DC range. (Zero adjustment is 

available only 400A DC range.)

(8) LCD  Display  :  Field  effect  type  of  liquid  crystal  display  with  maximum 

counts  of  3999  and  microprocessor-controlled  annunciators  and  the  deci-

mal point.

(9) Terminal Cover : Used to enclose the input terminals (Lo and Hi) when the 

OUTPUT terminal is in use, thus avoiding accidental application of voltage 

to the instrument.

(10) OUTPUT terminal (for current measurement only) : Provides DC voltage in 

proportion  to  the  reading  on  an  AC  or  DC  current  range.  The  voltage  is 

used  for  such  purposes  as  long  term  monitoring  with  a  recorder  or  other 

recording  devices.  This  terminal  cannot  be  accessed  on  a  voltage  or 

resistance range.

(11) Lo Terminal : Accepts the black test lead for voltage or resistance meas-

urement.

(12) Hi  Terminal  :  Accepts  the  red  test  lead  for  voltage  or  resistance  meas-

urement.

(13) Safety  Hand  Strap  :  Prevents  the  instrument  from  slipping  off  the  hand 

during use.

(14) Test Leads (Model 98072) : Connect to Lo and Hi terminals for voltage or 

resistance measurement.

(15) Output Plug (Model 98012) : Insert this plug into the OUTPUT terminal to 

obtain  DC  output  voltage.  Connect  suitable  connection  cord  to  the  plug 

when it is used. (See section 3.3 OUTPUT Terminal)

(16) Barrier:  It  is  a  part  providing  protection  against  electrical  shock  and 

ensuring the minimum required air and creepage distances.

Summary of Contents for CL250

Page 1: ...IM CL250 2017 10 11 KYOU IM CL250 User s Manual CL250 Clamp on Tester...

Page 2: ...ent 4 2 4 DC Voltage Measurement 5 2 5 AC Voltage Measurement 6 2 6 Resistance Measurement 7 2 7 Continuity Check 400 range fixed 7 2 8 MAX Measurement 8 3 Other Functions 9 3 1 Sleep Function 9 3 2 D...

Page 3: ...nual in order to avoid the risk of personal injury or death and or damage to the instrument Double Insulation This symbol indicates double insulation AC Voltage Current This symbol indicates AC voltag...

Page 4: ...hands are wet Do not exceed the maximum allowabIe input of any measurement range Never open the battery compartment cover when making measurement Do not use the instrument if there is any damage to t...

Page 5: ...ortable equipment etc with smaller transient measurement than CAT III Measurement category III CAT III Distribution level fixed installation with smaller transient measurement than CAT IV Measurement...

Page 6: ...into the output terminal Data Hold Switch operates as range selection switch See section 3 3 OUTPUT Terminal 5 Button Used to switch the instrument between the AC and DC modes The instrument is set to...

Page 7: ...Provides DC voltage in proportion to the reading on an AC or DC current range The voltage is used for such purposes as long term monitoring with a recorder or other recording devices This terminal ca...

Page 8: ...ement on a circuit above 1000VDC This may cause shock hazard Do not make current measurement with the test leads connected to the instrument Keep your fingers and hands behind the barrier during measu...

Page 9: ...may not reduce to nil even if the display is zero adjusted with the button In this case make zero adjustment on the recorder or other device that the output voltage is connected to Turing the Function...

Page 10: ...eading either The output voltage from the OUTPUT terminal may not reduce to nil even if the display is zero adjusted with the button In this case make zero adjustment on the recorder or other device t...

Page 11: ...ircuit above 750V AC This may cause electric shock hazard 1 Set the Function Selector switch to the 400V or 750V position If the instrument is in the DC mode press the button once to select AC mode Th...

Page 12: ...display NOTE When the tip of the test leads is shorted together the display may read a very small resistance instead of 0 This is the resistance of the test leads not fault If one of the test leads ha...

Page 13: ...This may cause electric shock hazard Do not make measurement with the test leads connected to the instru ment 1 Set the Function Selector switch to the desired position 2 Press the button to select th...

Page 14: ...enable the Sleep function turn the Function Selector switch back to OFF then to any other position NOTE The Sleep function is disabled while the output plug is inserted into the OUTPUT terminal When t...

Page 15: ...et the Function Selector switch to the 400A or 2000A posi tion The output is available only in the two ranges Proceed to meas urement in the DC or AC mode NOTE In the DC mode the output voltage from t...

Page 16: ...t range over 3000A but also clamp on a large bus bar or conductor 1 Set the Function Selector switch to the 400A position 2 Select the AC mode with the button 3 As shown in the figure below clamp MODE...

Page 17: ...n battery compartment If the instrument is powered on but the display blanks or is shown on the lower left corner of the display replace the batteries Note that when the battery is completely exhauste...

Page 18: ...Hz 5 0 rdg 4dgt 500 to 1kHz 2000A 0 to 1000A 1001 to 1999A 3 0 rdg 2dgt 50 60Hz DC Voltage V Input impedance 2M Range Measuring Ranges Accuracy 400A 0 to 399 9A 1 0 rdg 2dgt 1000A 0 to 999A AC Voltage...

Page 19: ...tinuity Check Display Liquid crystal display with maximum counts of 3999 Overrange Indication OL is displayed on each range Response Time Approx 2 seconds Sample Rate Approx 2 5 times per second Tempe...

Page 20: ...s Conductor Size Approx 55mm diameter max Dimensions Approx 105 W x 250 H x 49 D mm Weight Approx 530g with batteries Safety Standard EN 61010 1 EN 61010 2 033 EN 61010 031 EN 61010 2 032 AC DC 600V C...

Page 21: ...fails to operate correctly and needs repair contact the vendor from whom you purchased the instrument or the nearest Yokogawa Meters Instruments sales office Turn off the POWER switch once then turn i...

Page 22: ...IM CL250 17 This manual is valid only in China GB T 26572 GB T 26572...

Page 23: ......

Reviews: