IM CL250
3.3 OUTPUT Terminal (For current measurement only)
WARNING
● Do not make measurement on a circuit above 750VAC or 1000VDC. This
may cause electric shock hazard.
●Never apply voltage to the OUTPUT terminal.
(1) To obtain output voltage from the OUTPUT terminal, connect a suitable
cord to the supplied output plug.
(2) Slide the terminal cover to the right to enclose the Lo and Hi terminals.
Insert the output plug into the OUTPUT terminal for connection with a
recorder or other recording device.
(3) Set the Function Selector switch to the "
400A" or "
2000A" posi-
tion. (The output is available only in the two ranges.) Proceed to meas-
urement in the DC or AC mode.
NOTE
● In the DC mode, the output voltage from the OUTPUT terminal may not
reduce to nil even if the display is zero adjusted with the
button. In
this case, make zero adjustment on the recorder or other device that the
output voltage is connected to.
● The Sleep function is disabled while the output plug is inserted into the
OUTPUT terminal. When the output plug is disconnected from the terminal,
the Sleep function is enabled in about 10 minutes.
● Set the appropriate sensitivity on the recorder or other recording device.
(See chapter 5. Specifications.)