3-24
IM AQ6370E-02EN
Setting the Type of External Light Source and Calibration Wavelength Value
6.
Press
SYSTEM
. The SYSTEM menu is displayed.
7.
Tap the
Wavelength Calibration
.
Select the type of external light source (laser-type external light source or gas cell absorption line
type external light source) and set the calibration wavelength value.
There are two different ways to set the wavelength value.
For Laser Type Light Sources
8.
Tap the
External Laser
. A screen for specifying the wavelength of the external light source is
displayed.
9.
Select the laser wavelength value using the rotary knob or the arrow keys. The allowed
wavelength range is 600–1700 nm.
10.
Press
ENTER
. The wavelength value is set.
11.
Tap the
Execute
. Wavelength calibration is executed. The previous screen is displayed again
after the calibration process ends.
12.
Tap the
Cancel
during wavelength calibration to cancel the wavelength calibration process.
Enter the value.
Select from presets.
3.6 Wavelength Calibration