Preparing for Measurement
3-19
IM AQ6370E-02EN
1
2
3
4
5
6
App
Index
Alignment Adjustment
1.
Press the
SYSTEM
. The system menu is displayed.
2.
Tap the
Optical Alignment
.
• When A Stable Single-Wavelength Light Source Such as a DFB-LD Is Connected
3.
Tap the
External Laser
.
4.
Tap the
Execute
. Alignment adjustment is executed. The previous screen is displayed again
after the alignment adjustment ends.
• When the AQ6370E/AQ6370D Reference Light Source Is Connected
3.
Tap
Built-in Source
.
4.
Tap
Execute
. Alignment adjustment is executed. It will take approximately 2 minutes for the
alignment adjustment to complete. When the alignment adjustment finishes, the previous
screen returns.
5.
Tap the
Cancel
during alignment adjustment to cancel the alignment adjustment process.
3.5 Alignment Adjustment