5-2
IM AQ6150B-02EN
Item
Specifications
Functions
Measurement
Single, repeat, average, drift, data logging
Measurement
condition setup
Average count, air/vacuum wavelength, device type (CW/modulated), measurement range,
update rate switching
Display
Single wavelength, multi wavelength, delta, grid, spectrum (with zooming), wavelength axis
units (wavelength (nm)/frequency (THz)/wave number (cm
-1
)), optical power units (dBm/mW/
μW), center wavelength, total power, marker (up to 1024 points), label, power bar, warning
messages, error messages, system information
Data analysis
Peak search, FP-LD analysis, drift analysis, WDM (OSNR) analysis
File
Saving/loading measured results (CSV), saving/loading setup parameters (binary), saving
screen images (BMP), saving/loading logging data (binary and CSV)
Remote control
Interface selection (GP-IB/Ethernet), TCP/IP configuration, remote monitor
Others
Internal reference light source status LED, optical power offset, parameter initialization,
firmware updating
Interface
GP-IB, Ethernet, USB, VGA output
Remote control
GP-IB, Ethernet
Optical connector
FC/PC or SC/PC (AQ9441 Universal Adapter)
Warm-up time
60 minutes or more
Power requirements
100 to 240 VAC, 50/60 Hz, approx. 100 VA
Environmental Conditions
Optimal temperature: 10 to 30°C
Operating temperature: 5 to 35°C
Storage temperature: –10 to +50°C
Ambient
humidity: 20 to 85% RH (no condensation)
Dimensions and mass
Approx. 426 (W) × 132 (H) × 450 (D) mm (excluding protrusions), approx. 11 kg
Safety standards
Compliant standards
EN 61010-1 Pollution degree 2
9
Laser safety
EN 60825-1: 2014 Class 1
IEC 60825-1: 2007, GB 7247.1-2012 Class 1
Emissions
Compliant standards
EN 61326-1 Class A
EN 55011
Class A, Group 1
EMC Regulatory Arrangement in Australia and New Zealand EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard (
한국 전자파적합성기준
)
EN 61000-3-2
EN 61000-3-3
This is a class A instrument designed for an industrial environment. Operation of this
product in a residential area may cause radio interference in which case the user will be
required to correct the interference.
Cable conditions
• ETHERNET connector
Use Ethernet cables that are 30 m or less in length.
• VIDEO OUT connector
Use a shielded D-sub 15-pin VGA cable that is 3 m or less in length.
• USB interface
Use USB peripherals (i.e. mouse) with shielded USB cables that are 3 m in length or less.
• GP-IB connector
Use a shielded GP-IB cable that is 3 m in length or less.
9 Pollution degree refers to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand voltage or surface
resistivity. Pollution Degree 1 applies to sealed space (with no pollution or only dry non-conductive pollution). Pollution
Degree 2 applies to normal indoor atmospheres (with only non-conductive pollution).
5.1 Optical and General Specifications