4.3 Routine Maintenance
Cleaning the Ferrule End Face of the Optical Input
1.
Turn off the power. Then, remove the connector adapter from the
instrument
.
For the removal procedure, see section 2.4 in the Getting Started Guide, IM AQ6150-02EN.
2.
Clean the ferrule end face using a cotton swab moistened with a small amount of absolute
alcohol.
Always use a new cotton swab.
3.
When you finish cleaning, attach the connector adapter to the
instrument
.
Cotton swab
Optical input
Connector adapter
Optical connector
The instrument panel
Lock lever