IM 01E21A03-01EN

3

 

<2. Explosion Protection Type>

3

WARNING

 

AXG### magnetic flowmeter (Integral Flowmeter 

or Remote Sensor), and AXG4A Remote 

Transmitter are products which have been 

certified as explosion protection type products. 

Strict limitations are applied to the structures, 

installation locations, external wiring work, 

maintenance and repairs, etc. of these products. 

Sufficient care must be taken, as any violation of 

the limitations may cause dangerous situations.

Be sure to read this manual before handling the 

explosion protection type products.

WARNING

 

The terminal box cover and display cover are 

locked by special screw. In the case of opening 

the cover, use the hexagonal wrench (nominal 

size 3).

The covers of explosion protection type products 

are locked. Use this hexagonal wrench to open 

and close the cover. Before opening the cover, 

be sure to check that the power of flowmeter has 

been turned off. Once the cover is closed, be 

sure to re-lock the product.

Be sure to lock the cover with the special screw 

using the hexagonal wrench after tightening the 

cover.

2.1   Technical Data

Applicable Standard:

FM 3600, FM3610, FM3615, FM3616, FM3810, 

ANSI/UL UL 50E, NEMA 250, ANSI/UL 60079-0, 

ANSI/UL 60079-7, ANSI/UL 60079-11, ANSI/UL 

60079-18, ANSI/UL 61010-1, ANSI/UL 61010-2-30

Certificate:

FM 17US0140X

Integral Flowmeter 

Explosionproof (Explosion Protection Code: FF2)

•  Type of Protection:

 

CL I/II/III GP ABCDEFG  

IS CL I GP ABCD  

T6...T3

•  Enclosure: Type 4X

•  Ambient Temperature: 

 

See Table 2.1 and Table 2.2.

•  Process Temperature: 

 

See Table 2.1 and Table 2.2.

•  Power Supply:

 

100 to 240 Va.c. (50/60 Hz) / 100 to 120 Vd.c. 

24 Va.c. (50/60 Hz) / 24 Vd.c.

•  Um: 250 V

•  Current I/O: 4 to 20 mA, 32 Vd.c. max.

•  Digital I/O: 30 Vd.c. max., 200 mA max.

•  Fieldbus I/O: 32 Vd.c. max., 15 mA max.

•  Maximum Working Pressure: 

 

Read “Fluid Temperature and Pressure”

Explosionproof with Output IS Circuit 

(Explosion Protection Code: FJ2 and FT2)

•  Type of Protection:

 

CL I/II/III GP ABCDEFG  

IS CL I GP ABCD  

AIS CL I/II/III GP ABCDEFG 

T6...T3

•  Enclosure: Type 4X

•  Ambient Temperature:

 

See Table 2.1 and Table 2.2.

•  Process Temperature:

 

See Table 2.1 and Table 2.2.

•  Power Supply:

 

100 to 240 Va.c. (50/60 Hz) / 100 to 120 Vd.c. 

24 Va.c. (50/60 Hz) / 24 Vd.c.

•  Um: 250 V

•  Control Drawing: DIE016-A92

•  Maximum Working Pressure: 

 

Read “Fluid Temperature and Pressure”

2.  Explosion Protection Type

Summary of Contents for ADMAG TI Series

Page 1: ...ADMAG TI Series AXG Magnetic Flowmeter FM USA Explosion Protection Type IM 01E21A03 01EN 1st Edition IM 01E21A03 01EN User s Manual ...

Page 2: ... Edition 1st Edition Oct 2019 KP All right Reserved Copyright 2019 Yokogawa Electric Corporation Contents 1 Introduction 1 2 Explosion Protection Type 3 2 1 Technical Data 3 2 2 Cable Entry 7 2 3 Installation 7 2 4 Operation 12 2 5 Maintenance and Repair 12 2 6 Specific Condition of Use 12 Revision Information ...

Page 3: ...N A CAUTION sign denotes a hazard It calls attention to procedure practice condition or the like which if not correctly performed or adhered to could result in damage to or destruction of part or the entire product IMPORTANT An IMPORTANT sign denotes that attention is required to avoid damage to the product or system failure NOTE A NOTE sign denotes information necessary for essential understandin...

Page 4: ...neers responsible for installation and wiring of the instrument To ensure correct use read this manual and the applicable user s manuals as listed in IM 01E21A21 01Z1 Read Me First thoroughly before starting operation Read the general specifications as listed in IM 01E21A21 01Z1 Read Me First for its specification Trademark All the brands or names of Yokogawa Electric s products used in this manua...

Page 5: ... after tightening the cover 2 1 Technical Data Applicable Standard FM 3600 FM3610 FM3615 FM3616 FM3810 ANSI UL UL 50E NEMA 250 ANSI UL 60079 0 ANSI UL 60079 7 ANSI UL 60079 11 ANSI UL 60079 18 ANSI UL 61010 1 ANSI UL 61010 2 30 Certificate FM 17US0140X Integral Flowmeter Explosionproof Explosion Protection Code FF2 Type of Protection CL I II III GPABCDEFG IS CL I GPABCD T6 T3 Enclosure Type 4X Amb...

Page 6: ...Ambient Temperature 40 C to 60 C Temperature class T6 Power Supply 100 to 240 Va c 50 60 Hz 100 to 120 Vd c 24 Va c 50 60 Hz 24 Vd c Control Drawing DIE016 A92 Table 2 1 Temperature Ratings Model Size Process Connection Lining Temperature Table Integral Flowmeter Remote Sensor AXG002 AXG005 AXG010 AXG015 2 5 to 15 mm 0 1 to 0 5 in Wafer Flange PFA Lining Table A Table C AXG025 AXG032 AXG040 AXG050...

Page 7: ... C T3 T150 C 40 C to 150 C D 40 C to 45 C T6 T75 C 40 C to 50 C 40 C to 60 C T5 T90 C 40 C to 75 C T4 T120 C 40 C to 120 C T3 T150 C 40 C to 150 C E 10 C to 45 C T6 T75 C 10 C to 50 C 10 C to 60 C T5 T90 C 10 C to 75 C T4 T120 C 10 C to 120 C T3 T130 C 10 C to 130 C F 10 C to 45 C T6 T75 C 10 C to 50 C 10 C to 60 C T5 T90 C 10 C to 75 C T4 T120 C 10 C to 120 C T3 T150 C 10 C to 150 C G 10 C to 60 ...

Page 8: ... to 8 in flange type wafer type 250 to 300 mm 10 to 12 in flange type 350 to 400 mm 14 to 16 in flange type F0202 ai 1 1 For wafer types of 150 to 200 mm 6 to 8 in or carbon steel flange types process connection code C of 50 to 400 mm 2 to 16 in the minimum temperature is 10 C 14 F PFA Lining Remote Sensor and Integral Flowmeter high pressure type ASME Class 600 flange F0203 ai Pressure MPa psi Te...

Page 9: ...ring shall be in conduit as follows SEALALL CONDUITS WITHIN 18 INCHES In order to prevent the grounding conductor from loosening the conductor must be secured to the terminal tightening the screw with appropriate torque Care must be taken not to twist the conductor Warning In cases where the ambient temperature exceeds 50 C use external heat resistant wiring with a maximum allowable temperature of...

Page 10: ...IM 01E21A03 01EN 8 2 Explosion Protection Type 8 Read the Control Drawing All wiring shall comply with local installation requirements Integral Type Flowmeter Explosion Protection Code FJ2 F0213 ai ...

Page 11: ...IM 01E21A03 01EN 9 2 Explosion Protection Type 9 Remote Type Flowmeter Explosion Protection Code FJ2 F0214 ai ...

Page 12: ...IM 01E21A03 01EN 10 2 Explosion Protection Type 10 Integral Type Flowmeter Explosion Protection Code FT2 F0215 ai ...

Page 13: ...IM 01E21A03 01EN 11 2 Explosion Protection Type 11 Remote Type Flowmeter Explosion Protection Code FT2 F0216 ai ...

Page 14: ...orporation can repair the equipment 2 6 Specific Condition of Use In the case the electrodes and or grounding rings are made of titanium the sensor should be kept away from impacts and frictions in hazardous locations Electrostatic charges on the non metallic parts excluding glass parts or coated parts of the product shall be avoided The applicable temperature class ambient temperature range and p...

Page 15: ... 01E21A03 01EN 13 Revision Information Title ADMAG TI Series AXG Magnetic Flowmeter FM USA Explosion Protection Type Manual No IM 01E21A03 01EN Edition Date Page Revised Item 1st Oct 2019 New Publication ...

Reviews: