Specifications
Accessories : AA battery
×
2pcs., user’s manual, exclusive protective case
USB cable (Only for 53006)
Optional : Blackbody tape (Maximum available temperature 250
℃
)
Compliant standards.
IEC 60825-1
EN 60825-1
US 21 CFR 1040.10 and 1040.11
EN 61326-1
※
Specifications may change without prior notice.
Model
53005-E
53006-E
Measuring range
− 30 to 600℃
Field of view
□ 30/1000mm(D:S=33:1)
Optics
Si lens
Sensing element
Thermopile
Wavelength
8 to 14 μ m
Response time
0.5sec. / 90%
Accuracy(ε≒ 0.95)
− 30.0 to 0℃:± 3℃、0.1 to 200℃:± 2℃、
201 to 600℃:± 1%
Repeatability
± 1℃ of reading value
Display resolution
− 30.0 to 199.9℃:0.1℃、200 to 600℃:1℃
Sighting method
Coaxal laser marker(Class 2)
HOLD time
15 seconds
Continuous measurement mode
−
ON / OFF Selectable
USB output
−
○
Memory
1-point memory
35-point memory
High/Low Alarm LED/Buzzer
Alarm LED/Buzzer ON/OFF Selectable
Emissivity(ε)Adjustment
0.95/0.85/0.70 Selectable
ε Adjustable
(0.30 to 1.20/0.01step)
Display function
NOR / MAX / MIN
Power supply
AA Battery × 2pcs
Battery life
Approx. 15 Hours(With max load)
Ambient temperature
0 to 50 ℃
Ambient humidity
35% to 85%Rh(Without due condensation)
Storage temperature/humidity
− 10 to 60℃/35% to85% Rh
Material
ABS/TEEE
Dimension
H × W × D = Approx. 182 × 56 × 38mm
Weight
Approx. 250g(Including batteries)