IM 51011-01EN
39
KC
-
Yes
Selects the user area number of the color correction factor
Select the user area:KC,mm<delimiter>
m = 01 to 20
→ Response: KC,m<
delimiter
>
m = 0: successful, 1: error
KS
-
Yes Enables the color correction factor
Use the color correction factor: KS,m<delimiter>
m = 0: OFF, 1: ON
→ Response: KS,m<
delimiter
>
m = 0: successful, 1: error
LD
Yes
Yes
Selects how to send data from the lux meter to the PC
Select how to send: LD,m<delimiter>
m = 0: every measurement
1: every data request (RD) command (default)
→ Response: LD,n<
delimiter
>
n = 0: command successful, 1: command error
PO
Yes
Yes Enables automatic power-off
Enable automatic power-off PO,m<delimiter
>
m = 0: disable automatic power-off
1: enable automatic power-off
Note:
If this command is received while totalized intensity
of illumination or comparator measurement is in
progress, the command will be executed after the
meter returns to normal measurement.
→ Response: PO,m<
delimiter
>
m = 0: successful, 1: error