![YOKOGAWA 51011 User Manual Download Page 41](http://html.mh-extra.com/html/yokogawa/51011/51011_user-manual_911131041.webp)
IM 51011-01EN
40
RA
Yes
Yes
Sets auto range mode
Set auto range mode: RA<delimiter>
→ Response: RA,m<
delimiter
>
m = 0: successful, 1: error
RD
Yes
Yes Requests to send measured data
Request to send: RD<delimiter>
→
Response: RD,m<delimiter
>
m = 0: command successful, 1: command error
RH
Yes
Yes
Sets the range hold function
Set range hold: RH,m<delimiter>
m = 0: 9.99,* 1: 99.9, 2: 999,
3: 9,990, 4: 99,900, 5: 999,000
*: Model 51021 only
→ Response: RH,m<
delimiter
>
m = 0: successful, 1: error
RS
Yes
Yes Checks unsent (measured) data in buffer
Check: RS<delimiter>
→ Response: RS,m,n<
delimiter
>
m = 0: command successful, 1: command error
n = 0: no unsent data, 1: unsent data available
TM
Yes
Yes
Sets the transmitted delimiter
Set the delimiter: TM,m<delimiter>
m = 0: CRLF (default)
1: LF
2: 0x03
→ Response: 0<delimiter that was set before the
TM command was received
>