5.2 DETAILED OPERATING INSTRUCTIONS, CONT’D…..
5.22 OPERATION - MODE ADJUSTMENT, CONT’D…..
3.0 OUTSIDE Mode
NOTE: Ventilation Rate
Ventilate rate represents how much air is introduced into the indoor space from the
outside. Low values indicate low ventilation rates and potentially poor air quality.
High levels indicate excessive ventilation and potential excessive energy usage.
To obtain an accurate measurement, readings should be taken 2~3 hours after
occupancy has stabilized in a space or at a peak in daily CO2 concentrations
.
With indoor air quality control, CO2 value is an indicator of ventilation rate.
400 ppm
(Parts Per Million) is the default CO2 concentration outside
(according to
ASHRAE: American Society of Heating, Refrigeration and Air conditioning
Engineers).
3.1. Press the Mode button (
⑤),
until OUTSIDE flashes.
3.2. Press Enter (
②),
to show OUTSIDE CO2 and PPM flashes. Press Up/Down (
④/
③)
to adjust the reading.
3.3. Press the Enter (
②),
save the setting and return to the User Mode. After
modification, the VENT Rate will change.
4.0 CALI Mode
The CO2 monitor has been calibrated at the factory and should be recalibrated
every 12-months with a specified concentration of CO2. Sensor drift usually occurs
in linearity. To perform calibration, follow the steps below.
4.1. If battery icon shows low power, replace batteries before attempting to
calibrate or use AC adapter.
4.2. Press the Mode button (
⑤),
until CALI flashes.
4.3. Press Enter (
②),
“CALI” is shown on the display
4.4. Adjust the lower display to ambient CO2 value by pressing the Up/Down (
④/
③)
Button.
4.5. Press Mode button (
⑤)
for more than 10 seconds. “CALIBRATING” flashes.
Calibration will be finished after 5-minutes and the LCD will indicate
“PASS”. If LCD indicates “FAIL”, repeat the calibration.
4.6. Press Enter (
②),
and return to the User Mode