OXYGEN CONCENTRATOR USER MANUAL
5
CAUTIONS
(
Notice to the following content to avoid breaking
the machine
)
◆
Do not open the machine shell and internal boss to do the maintenance.
If users found some problems, or found abnormal alarm, do not
disassemble the repair, should contact the dealer or manufacturer
immediately.
◆
It should be placed in the clean and free of dust, on corrosive and
poisonous environment. Do not use this product under the condition of
strong magnetic field environment.
◆
The air entrance of the machine should be located in a well ventilated
and the air source inlet should be in place at least the pollutants
(including / pollutants refers to: burning gas, exhaust system, air vents
and other anesthesia vacuum exhaust port etc.).
◆
Please connect the separate (wall mounted) AC power outlet while using
this product, do not connect the Do not connect the row insert board
seat. Do not use with other electrical appliances at the same time use the
power supply socket.
◆
This machine is strictly prohibited contact with oil and grease, such as the
need to connect to other pipelines, valves, connectors, the use of this
product, the connection set before the installation should be cleaned to
ensure clean and no oil (grease) category. In the connection process, pay
attention to keep all parts of the cleaning, the connection of the whole
machine is strictly prohibited to touch any combustible oily liquid.
◆
To ensure that the bottom of the exhaust flow efficiency while use the
machine. The back do not paste the wall, at least 30 cm from the wall,
otherwise it will cause damage to the machine overheating.
◆
This machine is not frequent opening and closing, shut down after the
start, the time interval of not less than 5 minutes, so as not to affect the
service life of the compressor, the manufacturers recommend each
operation is better over 30 minutes.
◆
The machine is only used as a medical oxygen supply, the output of the