OXYGEN CONCENTRATOR USER MANUAL
22
Portable and mobile RF communications
equipment should be use no closer to any
part of the CP101 OXYGEN CONCENTRATOR
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommend ed separation distance
Conducte
d RF
IEC
61000-4-6
3
Vrms
150
kHz
to 80
MHz
3 V
d
=
P
]
V
5
.
3
[
1
d
=
P
]
E
5
.
3
[
1
80 MHz to 800 MHz
Conducte
d RF
IEC
61000-4-3
3
V/m
80
MHz
to
2.5
GHz
3 V/m
d
=
P
]
E
7
[
1
800 MHz to 2.5 GHz
Where p is the maximum output power rating
of the transmitter in watts
(
W
)
according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(
m
)
.b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey, ashould be less than the compliance
level in each frequency range. b
Interference may occur in the vicinity of
equipment marked with the following
symbol: