OXYGEN CONCENTRATOR USER MANUAL
21
yellow low Oxygen
light is on and the
intermittent audible
Signal is sounding.
2.
Air
intake
flitering foam is
blocked.
Check the Air intake flitering foam. If the flitering
foam is dirty, wash it following the cleaning
instructions on page 6.
3.
Exhaust
is
blocked.
Check the exhaust area: make sure there is nothing
restricting the unit exhaust.
F
Red
Service
Required light is on
and an intermittent
audible
signal
is
sounding.
1. Flow meter is
not properly yet.
Ensure the flow meter is properly set to the
prescribed.
2. Air
filter is
blocked.
To check if the Air filter have blocked, jam, please
clean up the clutter.
Check whether the Air filter is dirty, if dirty, please
timely replace as required on page 6.
3.
Exhaust
is
blocked.
Check the exhaust area: make sure there is nothing
restricting the unit exhaust. If the above remedies
do not work, contact your Medical provider.
If the machine is still not working properly, please contact the product supplier or service point.
* If there is no special description, please follow the above instructions.
ELECTROMAGNETIC IMMUNITY
Guidance and manufacturer’s declaration – electromagnetic immunity
The CP101 OXYGEN CONCENTRATOR is intended for use in the electromagnetic environment
specified below. The customer or the user of the CP101 OXYGEN CONCENTRATOR should assure
that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment-guidance