
OXYGEN CONCENTRATOR USER MANUAL
14
TECHNICAL PARAMETERS
Model
CP101
Working Voltage
230 V
~
± 10% / 50 Hz ± 1 Hz
Rated Power
(
VA
)
770
Output
Pressure=0/7kPa,
the flow range
(
L/min
)
1
~
10
Output
Pressure=0,the
purity value
(
within 10min
,
up to the standard
)
Oxygen flow=1
~
10 L/min
,
purity
≥
90%
Highest Flow
10 L/min
Highest Flow
with 7kPa
pressure
,
change of flow
≤
1 L/min
Highest Flow, purity
(
within
10min
,
up to the standard
)
≥
90%
Range of adusting flow
0
~
10L/min adusted continously
Net Weight
(
Kg
)
20.5
Noise dB(A)
≤
55
Dimension
(
mm
)
L400 × W300 × H540
Oxygen output pressure
60kPa-80kPa
Yellow low oxygen indicator
light
When 72%
≤
purity
≤
82%
,
the yellow indicator will
shin,contact the supplier, Make sure have the spare
oxygen nearby when use it continously.
Red alarm indicator light
When purity
≤
72%
,
red indicator will alarm
,
and
dispaly show Lo-o2.It will stop within one minute.Please
turn off at once and choose the spare oxygen , contact
the supplier