Operation Method
Apparatus/Liquid Sterilization Course
0B
★
In order to ensure the sterilization effect, the lower
exhaust valve knob should be turned to the micro
exhaust state (Figure 21) during the heating, pressure
rising and sterilization process to keep a small amount
of steam constantly discharged from the cylinder, so
that the temperature in the cylinder can become
uniform.
When the heater starts to work, both the
sterilization temperature lamp and the heating
lamp on the control panel light up, indicating
that the sterilizer has entered into the operation
state of heating, temperature and pressure
rising (Figure 19). When the temperature in the
cylinder rises to be about 102
℃
, the interlock
indicator lamp will turn to be normally ON
(Figure 20).
When the temperature of sterilizer rising to the
set temperature, the sterilizing lamp and the
sterilization time lamp on the control panel light
up, the working lamp turns to blink, and the
sterilization countdown appears in the display
window (Figure 22), indicating that the
sterilization articles are being sterilized inside
the cylinder of sterilizer.
The sterilization countdown is over, the word
"END" appears in the display window, the END
lamp on the control panel lights up and the
working lamp goes out, indicating that the
sterilization operation is completed (Figure 23).