16
Operation Method
Sterilizing Parameters
Use calibration offset function
Calibration offset is a function which corrects the difference between the temperature in furnace and
that of controller (sensor temperature) if arises. The function parallel corrects the difference either to
the plus or minus side within the whole temperature range of unit. The function can be set or
cancelled by the key.
①
Start operation with the target setting temperature. Check the temperature in furnace (temperature of
sample) with a thermograph after it is stabilized.
②
Check the difference between the setting temperature and that in furnace (temperature of sample).
③
Press the key. Select the character "SC", which indicates the calibration offset, using the "
▲▼
",
and then press the key.
④
Input the difference using the "
▲▼
" and then press the key. This completes the setting.
The setting range of offset correction temperature is +20
℃
to plus side and -20
℃
to minus side
respectively.
Current temperature
Corrected temp. to plus side
Corrected temp. to minus side