11
4. Operating Procedures
Performing shaking operation using the timer
About time setting
Timer setting time range is from 0 to 99.9 min. The minimum unit of
time for setting is 0.1 min (6 seconds
)
.
The timer volume may be turned either clockwise or anticlockwise for
setting.
About setting of a
shaking speed
The maximum shaking speed is 200r.p.m
(
times/min
)
.
Avoid setting to a speed higher than this, which causes severe
vibration of the unit and poses a danger.
About shaking stroke
The shaking stroke for this unit is fixed at 30mm.
When you want to
change the shaking
speed or the set time
You can change the shaking speed during operation simply by turning
the volume on the shaking speed adjusting panel.
You cannot change the set time during timer operation. Stop timer
operation first, and make setting again.
How to change the shaking system
The unit has employed a new system that allows selecting reciprocating, oval, or rotation shaking
system only with a volume.
Move the volume along the mark indicated on the top of the main unit.
When you fix the volume in the midpoint between two marks, the system will perform an intermediate
move between reciprocating and oval, or intermediate move between oval and rotation.
Reciprocating Oval Rotation
Shaking system
selector volume
Move the volume until it clicks and then tighten it at
that position.
MK161