14
4.
Operation Method
Operation Mode and Function List
All the operation modes of this unit are as below:
No.
Name
Description
Page
1
Fixed Temperature
Operation
Pressing the FIXED TEMP key enters into the fixed
temperature operation setting mode.
Pressing it again enters into the temperature setting mode.
The "▲▼" are used to set temperature.
Pressing the START/STOP key starts or stops operation.
P.18
2
Quick Auto Stop
Operation
This operation is used to specify the period up to automatic
stop during operation.
The period up to operation stop can be set by pressing the
TIMER key during fixed temperature operation.
The "▲▼" are used to set the time.
Pressing the START key starts the quick auto stop operation,
activates the timer function and stops the operation
automatically after specified period.
P.19
3
Auto Stop Operation
This operation is used to specify the automatic stop time in
the fixed temperature operation.
Pressing the TIMER key displays "AStp".
The setting temperature "SV" can be set by pressing the
ENTER key.
The operation time "tim" can be set by pressing it again.
Pressing the START/STOP key starts the auto stop operation.
P.21
4
Auto Start Operation
This operation is used to specify the period up to automatic
start after power on.
Pressing the TIMER key displays "AStr".
The setting temperature "SV" can be set by pressing the
ENTER key.
The operation time "tim" can be set by pressing it again.
Pressing the START/STOP key starts the auto start operation.
P.23
5
Program Operation
This operation is used to change the temperature according to
the setting temperature and time.
Pressing the PROGRAM key displays "PrG1".
Press it again to select the program mode.
Press the ENTER key to select the pattern "PA t". Press the
ENTER key to display "End". Input the number of patterns to
be used.
Input the temperature and time of patterns "SV-n" and "t-n"
respectively.
P.25
※
This unit is impossible to change mode during the operation. If the mode requires to be
changed, stop the operation.