25
4.
Operation Method
Program Operation
Program
operation
This operation is used to change the temperature according to the setting
temperature and time.
Temp.
△
Start
△
Stop
Time
Program types
Six patterns of program types maximum can be input.
PrG1
―
1 program pattern using 30 steps maximum can
be created.
PrG2
PA
t
1
2 program patterns using 15 steps maximum can
be created.
PA
t
2
PrG3
PA
t
1
3 program patterns using 10 steps maximum can
be created.
PA
t
2
PA
t
3
Before
inputting
program
Input program patterns before program operation.
①
Check the number of steps in a created program and their setting
temperature/time. Refer to page 25~27.
②
Check the temperature rise/fall capability of the unit. Set the time
within the capability above. Suppose, for instance, that in the unit
which has capability of increasing or decreasing temperature by
50
℃
within 15 minutes, about 30 minutes is needed to increase or
decrease temperature by 100
℃
from current temperature
Repeat function
Repeat function is useful in case the operation uses the program
repeating the same program steps. Refer to page29 for the
function.
③
Check if the controller has sufficient free pattern for the number of
steps to be created. The steps, however, using the repeat function
mentioned above are not counted.