13
How to cancel addition
The DP-5601 has a function to cancel the added data in the scale memory.
There are two ways to cancel added data and the way can be chosen by parameter setting
#086. The default setting is “”086 = 0”.
Setting value
Name
Function
Remarks
0
Cancel last addition
Cancel last minute addition data
Only once available.
1 Subtraction
of
weight
Subtract the weight on platform from
the scale memory.
Multiple subtractions
available.
1. How to cancel last addition
Operating procedure
Panel display
Remarks
c
Press the
key after addition.
If the parameter number 035 is set to
any numbers other than “0,” “SEnd”
appears for the preset seconds.
The last addition
weight is canceled,
regardless of the
displayed weight.
2. How to subtract weight
Operating procedure
Panel display
Remarks
c
Place weighing items you desire
to subtract on platform after
addition.
The display shows
the weight of items.
d
Press the
key.
The display remains the same.
If the parameter number 035 is set to
any numbers other than “0,” “SEnd”
appears for the preset seconds.
e
Remove the items from the
platform.
The display returns
to “0.00.”
f
- One cycle of subtraction of weight is now complete.
- By repeating Steps
d
to
e
, the continuous subtractions of weight from the internal memory are
available.