17
Operation Method
Auto Stop Operation
As shown in the following figure, the device stops operating automatically by setting the timer.
Setting of the Auto Stop Operation
1
Turn on the power
POWER
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FAN
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
SUBMENU
POWER
TROUBLE
℃
Standby
2002/01/01 12:00
DOOR
•
Turn on the power switch of the unit (earth leakage
breaker). Pressing the POWER key turns on the
power. The Main Display indicates the
temperature in furnace. The Sub Display
indicates "Standby" and the STANDBY lamp lights
on. (Hereafter, this state is called the "standby
state".)
•
The FAN lamp lights on and the fan rotates when
"Standby" is displayed. Refer to the page 50 to
stop the rotation of fan.
2
Select operation mode
MENU
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FAN
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
SUBMENU
POWER
TROUBLE
℃
Auto-stop OPR
Select OPR mode
DOOR
①
Pressing the MENU key displays the operation
mode selection screen.
②
The Sub Display on the operation mode selection
screen displays the name of operation mode
currently selected with blinking. The
corresponding operation mode lamp blinks at the
same time.
③
Keep pressing the MENU key until the auto stop
operation mode is displayed.
④
Press the ENTER key. The auto stop operation
mode is decided.
The fixed temperature operation is selected at the
initial setting of unit. The operation mode carried
out last is selected in case other than it.
Operation
period
SV:
Setting temperature
t:
Time
t
Set timer
▼
SV
Activate timer
Start operation (manual) Stop
operation
(auto)
Waiting