Operation Method
Program Creation Method
1
Edit step
•
The unit operates with fixed rate inclination if the "START TEMP" and the "Set Temp" are different.
It enters into "Wait" if the temperature does not reach to the setting temperature within a setting
time.
Set Temp.
TEMP
.
TIME
Set Time
TEMP
.
START
.
(Operation with fixed rate inclination)
•
The temperature is kept constant till the end of setting time when the "START TEMP" and the "Set
Temp" are the same. The unit enters into "Wait" and measurement of residual time is suspended
when the temperature in furnace is 6
℃
over or 3
℃
below to the setting temperature.
Set Temp.
TEMP.
TIME
Set Time
TEMP
.
START
.
•
The unit operates with full-power from the "START TEMP" to the " Set Temp". It keeps the "Wait"
state until the temperature in furnace comes over 6
℃
or below 3
℃
to the setting temperature.
Set Temp.
TEMP.
TIME
Reached time to the set temp.
TEMP
.
START
.
Waiting