15
Operation Method
Fixed Temperature Operation
Start the operation from turning on the power shown in the figure, and continue the operation
under the setting temperature unless turning off the power.
Setting of the Fixed Temperature Operation
1
Turn on the power
POWER
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FAN
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
SUBMENU
POWER
TROUBLE
℃
Standby
2002/01/01 12:00
DOOR
•
Turn on the power switch of the unit (earth leakage
breaker). Pressing the POWER key turns on the
power. The Main Display indicates the
temperature in furnace. The Sub Display
indicates "Standby" and the STANDBY lamp lights
on. (Hereafter, this state is called the "standby
state".)
•
The FAN lamp lights on and the fan rotates when
"Standby" is displayed. Refer to the page 50 to
stop the rotation of fan.
2
Select operation mode
MENU
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FAN
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
SUBMENU
POWER
TROUBLE
℃
Fixed TEMP OPR
Select OPR mode
DOOR
①
Pressing the MENU key displays the operation
mode selection screen.
②
The Sub Display on the operation mode selection
screen displays the name of operation mode
currently selected with blinking. The
corresponding operation mode lamp blinks at the
same time.
③
Keep pressing the MENU key until the fixed
temperature operation mode is displayed.
④
Press the ENTER key. The fixed temperature
operation mode is decided.
The fixed temperature operation is selected at the
initial setting of unit. The operation mode carried
out last is selected in case other than it.
SV
t
▲
START
▲
STOP
SV:
Setting temperature
t:
Time