31
PROGRAMMED OPERATION METHOD
TO MAKE OUT NEW PROGRAM
・
First of all, make sure that the number of remaining segment is not 0. In case the
number is 0, delete either registered program before making new one.
1.
To choose “Prepare Program”
・
On function choice screen in program menu, have “Prepare Program” display and press
jog dial.
Prepare Program
Entry 0 rest 16
℃
FIXED TEMP
AUTO START
AUTO STOP゚
PROGRAM
2.
To choose “Program number”
・
Program number choice screen appears, sub display indicates “Prepare” and the number
which indicates program number blinks. Turn jog dial to have the program number
which you need display from “P01
~
P99”.
*
However program number which is already registered is not displayed.
Prepare
P01
Entry
0
rest
16
(Program number choice screen)
3.
Press “Jog dial”
Prepare P01
Entry 0 rest 16
℃
FIXED TEMP
AUTO START
AUTO STOP
PROGRAM
・
This fixes program number. When program number is decided, “Edit Program” screen
appears, so that you proceed to page 33.
Summary of Contents for DF Series
Page 2: ......
Page 62: ...59 SCHEMATICS DF811 ...
Page 63: ...60 DF1011 ...
Page 64: ...61 DH811 ...
Page 65: ...62 DH1011 ...
Page 66: ...63 ...
Page 70: ...67 ...
Page 71: ......