5
2. PRE-OPERATION PROCEDURES
Installation Precautions & Preparations
WARNING!
1. Choose an appropriate installation site.
●
Do not install unit:
・
where installation surface is not
completely level, not even or not clean.
・
in excessively humid or dusty locations.
・
where power supply is erratic.
・
where ventilation is not sufficient
・
where flammable or corrosive
gases/fumes will be generated.
・
where external temperature will exceed
30ºC, will fall below 5ºC or will fluctuate.
・
where there is constant vibration.
・
in direct sunlight or outdoors
Install unit in a location with sufficient space as specified below.
More than
20cm
More than
20cm
Front side
More than 40cm
2. Install unit in a safe, well-ventilated area.
In the event of an earthquake or other unforeseen incident, equipment may unexpectedly shift or
fall, causing injury. Taking preventative steps to install unit in a safe location, away from room
access doors and out of other danger is strongly recommended.
This unit is equipped with an air-cooled refrigerator. Install unit in a well-ventilated or
air-conditioned place so that unit is allowed to sufficiently diffuse heat. Failure to do so may
cause ambient temperature to rise, resulting in possible degraded cooling performance, overload
on the refrigerator, or equipment malfunction.
3. Install in a location free of flammables and explosives.
●
Never install near flammables or
explosives. This unit is NOT fire or blast
resistant. Simply switching the main
power switch (ELB) “ON” or “OFF” can
produce a spark, which can relay during
operation, causing a fire or explosion
when near flammable or explosive
fluids, chemicals or gases/fumes.
Refer to “List of Hazardous Substances”
(P.29).
Main
Unit
Explosives
Flammables