10. SPECIFICATIONS
Main Unit
Model DC801
Method
Cooled container system
Performanc
e
Drying capacity
Max. 1.0L
Minimum temperature *2,3
-85°C or less
Configuration
Refrigeration
Air cooling, 350W
Refrigerant Isobutane,
Ethylene, Methane
Temperature display *4
7-segment LCD
Temperature sensor
Platinum resistance temperature detector
Pt100
Ω
Pressure meter
Pirani gauge
Defrost mechanism
Hot gas bypass type
Vacuum port
φ
17mm
Chamber material
SUS304, Cylindrical
Dimensions of chamber
φ
153×H235mm
Capacity Approx.
4L
Usable external temperature
5 to 30°C
Safety
Safety devices
Over-current (20A) earth leakage breaker, refrigeration system
overload relay, high-voltage refrigeration error, refrigeration delay
timer, sensor failure error
S
tandard
measurement
s
External dimensions *5
W405×D500×H846 mm
Weight Approx.
65Kg
Power requirements/capacity
*6
AC100V±5
%
7/6.5A (20A)
Service outlet capacity AC100V Total 8A
Included items
Vacuum silicon grease (TORAY H.V.G), Vacuum hose (
φ
12 x
φ
30 x
1.5m), Instruction manual, Pirani gauge instruction manual,
Guaranty card, Fuse (10A), Caster stop x 4
*1. Performance based on operation at 23°C±5°C, 65%RH±20% humidity, AC100V/50Hz rating, no load.
*2. Measured at 140mm from the top of trap wall seal.
*3. Value where the cold trap is used as an air reservoir
*4. Display temperature is a surface temperature of refrigeration piping and served as a guideline.
*5. Dimensions do not include protrusions.
*6. The current value in brackets represents breaker capacity