Dental X-Ray Unit
Z70
70kVp - 7mA
4
2. Technical Data
Operation Manual
1.6. Use Warning
This X-Ray equipment produces ionization radiation which might damage healthy when operated
incorrectly. Only well trained professional can operate the equipment according this Manual and
current legislation. Be careful during the Arms and Tubehead operating. The rotative moviments
can hurt the hands and/or fingers.
- Even if compliant to specifications of electromagnetic compatibility, it is recommended not
to use the equipment in presence of external electromagnetic fields, such as those generated
by cellular phones, which might interfere with the electronic circuits of the system.
- The User takes legal responsibility on the possession, installation and use of the equipment
- The equipment is not designed to be used in the presence of flammable anesthetics, oxygen or
nitrous oxide
- To protect the patient from X-Ray, radiation protection accessories, such as standard leaded
aprons must be used
A front leaded apron should cover the patient down to mid-thigh. A thyroid apron should also
be used. For children, appropriately sized shielding devices should be used, including a front
apron and an adjustable thyroid/neck protector
- Follow all the instructions of this Manual. XZeal Technologies, Inc. is not responsible
for any injury caused by badly or inadequate use of this product
- Voltage
VAC
: 120 or 230
±10% - Single phase v ground
- Frequency
/60
Hz
: 50
- Electric current 8
4A
: A (120V) or
(230V)
- Max power consumption 0.9 kW
:
- Fuse:
6.3 A
10A (120V) or
(230V) Line and 500 mA board protection
- Focus: 0. m
(IEC 60336)
7 m X 0,7mm
- Anodic X-Ray tube angle:
16°
- Tube voltage 70 V
10%
:
k p ±
- Anodic electric current
mA 15%
: 7
±
-
cycle
/
3 sec
Duty
: 1 0
- Total Filtration
1 mAL
/
:
≥
2, m
Equiv./70kV
- Inherent Filtration
0.5mmAL
:
≥
- HVL: 1.5mmAL/70kV
≥
- Leaking radiation: <0.25mGy/[email protected] (measured 70kV/7mA/2.5sec exposition)
- X-Ray tube model: TOSHIBA D-0712
- Beam Limit Device SSD (Skin source distance): 22cm (8.5/8
’’
)
- Output radiation field (diameter):
6cm (2.3/8"
)
≤
Ø
2.1. Technical Specifications