Warning
Normally the P1 and P2 failure means the using power voltage
is not stable .We suggest the
installation of a 1.5KW automatic AC Voltage Regulators.
Code
Description
Signal
Operation
Solution
P1
The voltage of power is lower than
exposure voltage
Red alarm lamp is on/The
LED display error code P1
Prohibit any
operation
Waiting for automatic restoration
after voltage restoration
P2
The voltage of power is higher than
exposure voltage
Red alarm lamp is on/The
LED display error code P2
Prohibit any
operation
Waiting for automatic restoration
after voltage restoration
P3
The exposure finished and the tube is in
state of cooling, pausing the functions
The LED display code P3
Prohibit
exposure
Waiting for automatic restoration
after pre-set time (1/60)
P4
The exposure process ended earlier than
adjusted
The LED display code P4
Prohibit
exposure
Waiting for automatic restoration
P5
Line released between controller and
Tubehead or Tubehead broken
The LED display code P5
Prohibit
exposure
Contact a specialized professional
or XZeal to solutions