Dental X-Ray Unit
Z70
70kVp - 7mA
9
C. Protection Against Explosion
The X-Ray equipment must not be used in the presence of desinfectants, flammable or potentially
explosive gases and vapors that might catch fire and cause damage
In case these disinfectants have to be used, let the vapors completely disperse before turning on
the X-Ray equipment
D. Mechanical Risk
- After the installation, check with the technician the equipment assembly and support, respecting
all specifications of this Manual
- The internal arm spring is under high pressure. Be carefull when handling the arms to avoid any
injurie
- Make sure the equipment during the adjust position won´t nip the finger from the patient or
operator
E. Protection Against Radiation
- The local of equipment installation must be in compliance with the local standards, for the
operator, patient and othe persons protection against radiation
- The Z70 unit must be used in compliance with the local standards concerning radiation
protection
- During the examination, all personnel must comply with safety regulations concerning protection
against radiation. For the operator safety, keep at least 2 meters (6.5 fts) from the X-Ray source
- Only qualified personnel and for the intended use must operate the equipment
- The operator must be responsible for the patient protection against unnecessary or excessive
radiation doses
- Protection devices as apron and thyroid collar are required to protect the patient from radiation
3.4. Disposal - Enviroment
For disposal the equipment, contact the local authorities for informations. Proper disposal of this
product will help protect the enviroment
3.5. Equipment Cleaning
- Turn off and disconnect the equipment from the main electrical power supply before cleaning
- Do not spray any liquid/product directly on the device
- The Z70 is not waterproof. Use a soft fabric with the cleaning not saturated solution
3.6. Equipment Infection Control and Desinfection
- Infection Control - Protect the system from contamination using barriers membrane/sleeves/
sheets on the parts with operator and/or patient contact, available from TotalCare™ or Dental
Distributors
- Disinfection - The equipment can be cleaned and/or disinfected using the solution [Disobutyl
Phenoxethoxycthil Dimethyl Benzyl Ammonium Chloride]
NOTE: Do not use solvent or caustic disinfect
3.7. Equipment Maintenance
The equipment requires proper operation, periodic maintenance and servicing. The following
precautions will ensure safe and effective functioning of the system.
Periodic maintenance consists of system checks directly performed by the operator and/or by
Technical Service.
Operation Manual