background image

14

15

Next, use the up or down arrow keys to adjust the shut off time from 0 
to 60 minutes. Press enter to save the new shutoff time. Next, highlight 
the ESC-Exit box and press enter to close the submenu. 

To disable Auto Shutoff, set the Time in Minutes to 0 (zero).

cell constant

The Cell Constant displays the cell constant of the conductivity cell. 
The cell constant is calculated and updated each time a conductivity 
calibration is performed. The cell constant range is 4.0 to 6.0. Resetting 
the System Menu resets the cell constant to 5.0.

resettIng the systeM setUp MenU and 

cell constant to factory defaUlt

To reset the Pro1030 settings and conductivity cell constant back to 
factory default, press the down arrow key while in the System Setup 
menu until the Reset -   box is highlighted, then press enter.  The 
instrument will prompt you to confirm the reset.  Highlight Yes and 
press enter to continue with the reset or highlight No and press enter 
to cancel the reset.  A Factory Reset will not affect data saved in the 
instrument’s memory.  

The following will be set in the Pro1030 after performing a reset:

Parameter

Reset Defaults

Audio

on

Contrast

Set to mid range

Temperature Units

°C

ISE Sensor Type

pH

ISE Units

pH

ISE Auto Stable

off (0.0 % Change 

and 10 seconds)

pH Buffer Set

USA

Conductivity Units

cond mS/cm and spc mS/cm

Parameter

Reset Defaults

Conductivity Auto Stable

off (0.0 % Change 

and 10 seconds)

SPC Reference Temperature

25°C

SPC Temperature Coefficient

1.91%/°C

TDS Constant

0.65

Language

English

Auto Shutoff

30 minutes

Conductivity Cell Constant

5.0

pH Calibration

Factory default

C A L I B R A T I o N 

teMperatUre 

All Pro1030 cables have built-in temperature sensors. Temperature 
calibration is not required nor is it available.

ph calIBratIon

The Pro1030 pH sensor can be calibrated by performing a 1, 2 or 3-point 
calibration. At least one of the calibration points must be done with 
pH buffer 7 or 6.86. For auto buffer recognition to work properly with 
an older or dirty sensor, calibrate in buffer 7 or 6.86 first. For highest 
accuracy, use fresh, traceable pH buffers and ensure the sensor and 
calibration vessel are clean.

1-PoINT CALIBRATIoN

1.  Place the sensor in pH buffer 7 or 6.86 and allow the temperature 

and pH readings to stabilize. 

2.  Press and hold Cal 

 for three seconds. 

3.  Highlight pH and press enter. If pH is not listed as an option, 

check the System Setup menu to ensure pH is enabled in the 
ISE Sensor Type menu.

4.  Highlight 1 point and press enter.
5.  If necessary, use the up and down arrow keys to adjust the pH 

buffer value. Note the pH mV reading which ideally should be 
between -50 and +50 in buffer 7.

6.  Press enter to complete the calibration or press Cal 

 to cancel. 

Summary of Contents for YSI Pro1030

Page 1: ...USER MANUAL English Pro1030...

Page 2: ...First Power On 7 System Setup Menu 7 Audio 8 Contrast 8 Temperature Units 8 ISE Sensor Type 9 ISE Units 9 Auto Stable 9 pH Buffer Set 10 Conductivity Units Cond Units 10 Specific Conductance Reference...

Page 3: ...red by this warranty To exercise this warranty call your local YSI representative or contact YSI Customer Service in Yellow Springs Ohio at 1 937 767 7241 800 897 4151 or visit www YSI com for a Produ...

Page 4: ...pack the instrument and accessories and inspect for damage Compare received parts with items on the packing list If any parts or materials are damaged or missing contact YSI Customer Service at 800 89...

Page 5: ...u from the Run screen 2 Up Arrow Use to navigate through menus to navigate through box options along the bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to t...

Page 6: ...y 1 o ring visible Failure to properly seat the sensor may result in damage 6 Twist the sensor clockwise to engage the threads and finger tighten Do NOT use a tool This connection is water tight The I...

Page 7: ...System Setup menu the up and down arrow keys will move the highlighted bar up and down the system setup options See the System Setup menu section of this manual for more information about these option...

Page 8: ...ther pH model 1001 or ORP model 1002 Use the up or down arrow keys to highlight ISE Sensor Type then press enter to open a submenu Highlight the sensor type corresponding to the sensor installed on th...

Page 9: ...enter to enable or disable it An enabled conductivity unit will have an X in the box next to it Highlight the ESC Exit box along the bottom of the display and press enter to save any changes and to cl...

Page 10: ...e a multiplier for the water at your sampling site Use the following procedure to determine the multiplier for a specific sample 1 Determine the specific conductance of a water sample from the site 2...

Page 11: ...e and 10 seconds pH Buffer Set USA Conductivity Units cond mS cm and spc mS cm Parameter Reset Defaults Conductivity Auto Stable Off 0 0 Change and 10 seconds SPC Reference Temperature 25 C SPC Temper...

Page 12: ...hold Cal for three seconds 3 Highlight pH and press enter If pH is not listed as an option check the System Setup menu to ensure pH is enabled in the ISE Sensor Type menu 4 Highlight 3 point and pres...

Page 13: ...hloride when calibrating Specific Conductance since these solutions typically have a Temperature Coefficient of 1 91 C Additionally be sure to enter the value of the solution as it is listed for 25 C...

Page 14: ...ess the Cal key to exit the calibration error message and return to the Run screen See the Troubleshooting section of this manual for possible solutions Taking M easurements Before taking measurements...

Page 15: ...the data set number 01 through 50 Erasing Data While viewing saved data press the enter key to access the function boxes at the bottom of the display Next use the up or down arrow keys to highlight E...

Page 16: ...prevent water from entering the port Once the ISE sensor is removed examine the connector inside the port If any moisture is present use compressed air to completely dry the connector or let it air d...

Page 17: ...h all parts of the sensor tip without stress You can also use a pipe cleaner for this operation if more convenient i Dry the port and sensor connector with compressed air and apply a very thin coat of...

Page 18: ...displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Ensure the conductivity sensor is clean Follow the cleaning procedures in the Care Maintenance and Storage section...

Page 19: ...uctivity 0 500 uS cm 0 5 mS cm 0 50 mS cm 0 200 mS cm auto ranging 0 0001 to 0 1 mS cm 0 1 to 0 uS cm range dependent Instrument only 0 5 of the reading or 1 uS cm whichever is greater Instrument with...

Reviews: