Technical data
UV 70x IQ SAC
46
ba77046e02
05/2017
8.2
Application characteristics
Suitability and areas
of application
Measuring medium
Pressure resistance Sensor with connected SACIQ (SW) sensor connection cable:
The sensor meets all requirements according to article 3(3) of 97/23/EC ("pres
-
sure equipment directive").
Type of protection Sensor with connected SACIQ (SW) sensor connection cable:
IP X8, 1 bar (10
5
Pa)
Immersion depth
BOD correl.
0.0 ... 500.0 mg/L
0.1 mg/L
DOC correl.
0.0 ... 500.0 mg/L
0.1 mg/L
UVT 254, dissolved
*
0.0 ... 100.0 %
0.1 %
* The UVT-254 value is standardized to 10 mm gap width.
The quoted measuring ranges are nominal measuring ranges that
are theoretically possible. In practice, real measuring ranges exist
that are given by the limits of photometric determination. The limits
are significantly influenced by the light scattering due to solids and
the absorption of accompanying substances (sample matrix).
Therefore, no fixed measuring range can be determined before
-
hand.
Measuring mode
Measuring range
Resolution
Municipal wastewater with a low proportion of industrial wastewater, waste
water treatment plants, surface water
pH value
4 - 12
Flow rate
< 3 m/s
Temperature
0 °C ... + 45 °C
(32 °F ... + 113 °F)
Max. admissible overpressure
105 Pa (1 bar)
min
The measuring gap must be
filled with test sample.
max
10 m depth