UV 70x IQ SAC
Commissioning
21
ba77046e02
05/2017
3.4.4 Settings for virtual sensors
Carrying out
settings
Using <S>, switch from the measured value display to the main menu of the
settings. Then navigate to the setting menu (setting table) of the sensor. The
procedure is described in detail in your IQ S
ENSOR
N
ET
system operating
manual.
Default values are marked in bold.
Quit
-
The display switches to the next higher
level without storing the new settings.
Menu item
Settings
Explanations
Menu item
Settings
Explanations
Measuring mode
SAC 254, dissolved
SAC 254, total
UVT 254, total
CODds correl.
CODto correl.
TOC correl.
BOD correl.
DOC correl.
UVT 254, dissolved
The measured parameter is displayed
in the selected citation form.
Measuring range
-
Display of the measuring range
(see chapter 8 T
).
Cal - # raw value 1
Value pairs of the user calibration (see section 4.3.2).
Note: Leave those values unchanged during the initial commission
-
ing.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Main sensor
Information that there are further sen
-
sor overlapping settings in the setting
menu of the relevant main sensor
(
Meas. interval
,
Cleaning duration Air
,
Signal smoothing
etc.).
Save and quit
-
The system confirms the saving of the
settings and the display switches to the
next higher level.