OPERATING MANUAL
ba77046e02 05/2017
UV 701 IQ SAC
UV 705 IQ SAC
OPTICAL SAC SENSOR
Page 1: ...OPERATING MANUAL ba77046e02 05 2017 UV 701 IQ SAC UV 705 IQ SAC OPTICAL SAC SENSOR...
Page 2: ...2 ba77046e02 05 2017 UV 70x IQ SAC Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 3: ...2 Requirements for safe operation 9 2 2 3 Unauthorized use 9 3 Commissioning 10 3 1 IQ SENSOR NET system requirements 10 3 2 Scope of delivery of the UV 70x IQ SAC 10 3 3 Installation 11 3 3 1 Mounti...
Page 4: ...cleaning 38 5 2 4 Cleaning the measuring gap 39 6 Spare parts maintenance equipment accessories 40 7 What to do if 42 8 Technical data 45 8 1 Measurement characteristics 45 8 2 Application characteris...
Page 5: ...g manual has a modular structure like the IQ SENSOR NET itself It consists of a system operating manual and the operating manuals of all the components used Please file this component operating manual...
Page 6: ...l cases and thus ensures reliable measured values during continuous operation Compressed air cleaning option The sensor is designed to be equipped with an additional compressed air driven cleaning sys...
Page 7: ...the reference beam 8 over a distance without measuring solution The test sample is between both measurement windows 3 in the measuring gap The receiver of the optical system 4 directs the measuring b...
Page 8: ...ght be damaged if the actions mentioned are not taken 2 1 2 Safety signs on the product Note all labels information signs and safety symbols on the product A warning symbol triangle without text refer...
Page 9: ...for safe operation Note the following points for safe operation The product may only be operated according to the authorized use specified above The product may only be supplied with power by the ener...
Page 10: ...f delivery SensorUV 70x IQ SAC Sensor sleeve Shock protectors Operating manual DIQ S 28X Controller software Version 3 72 or higher DIQ S 182 Controller software Version 3 45 or higher MIQ MC3 Control...
Page 11: ...t contact corrosion NOTE Do not suspend the sensor on the sensor connection cable Risk of cable break and water penetration at the cable gland A wide range of mounting accessories is available for the...
Page 12: ...te is 0 2 m s so the measuring gap is kept clean optimally Fig 3 2 Vertical installation Protection against shocks When installing the sensor always make sure it cannot knock against a wall or another...
Page 13: ...tors proceed as follows Fig 3 4 Mounting the shock protectors 1 Plug the cap pos 2 in Fig 3 4 on one of the two rings pos 1 2 Put the assembled unit on the sensor end and fix it with 2 cable ties pos...
Page 14: ...ET accessories is given in the WTW catalog and on the Internet Are the plug connections dry Before connecting the sensor and sensor connection cable please make sure the plug connections are dry If mo...
Page 15: ...and the SACIQ SW sensor connection cable and keep them safe 2 Plug the socket of the SACIQ SW sensor connection cable onto the plug head connector of the sensor At the same time rotate the socket so t...
Page 16: ...asure After the initial commissioning a user calibration is required for all correlated parameters CODds correl CODto correl TOC correl BOD correl DOC correl see section 4 3 CAUTION Never look into th...
Page 17: ...ion UV 70x IQ SAC and the series number of the physical sensor Virtual sensors All other sensors for carbon measurement are virtual sensors You register on the system under the designation UV 70x IQ S...
Page 18: ...witch from the measured value display to the main menu of the settings Then navigate to the setting menu setting table of the sensor The procedure is described in detail in your IQ SENSOR NET system o...
Page 19: ...e selected measuring mode the measurement location is automatically changed Measuring range Display of the measuring range see chapter 8 TECHNICAL DATA Cal raw value 1 Value pairs of the user calibrat...
Page 20: ...the automatic time controlled cleaning function Cleaning type Ultrasound Air Ultrasound air Cleaning methods adjustment time This menu item is only available with Cleaning mode On The duration of the...
Page 21: ...dissolved SAC 254 total UVT 254 total CODds correl CODto correl TOC correl BOD correl DOC correl UVT 254 dissolved The measured parameter is displayed in the selected citation form Measuring range Di...
Page 22: ...Commissioning UV 70x IQ SAC 22 ba77046e02 05 2017 Quit The display switches to the next higher level without storing the new settings Menu item Settings Explanations...
Page 23: ...ig 4 1 Determination of measured values During a measurement the spectrum of the test sample is recorded and set against the stored absorbance of a zero adjustment measurement in ultrapure water The r...
Page 24: ...windows due to difficult sample matrix Damaged measuring windows Lacking quality of the zero adjustment e g due to contaminated ultrapure water The user calibration is not representative or erroneous...
Page 25: ...tamination or mechanical damage in the area of the measuring chamber or on the measurement windows affect the measured values A Sensor check H2O dest determines the condition of the sensor By checking...
Page 26: ...sor was thoroughly cleaned If the Zero adjustment is not successful the sensor is blocked for measurement A successful Zero adjustment is required for the determination of measured val ues Sensor chec...
Page 27: ...point of the applica tion This method results in a single point calibration If the application has to cover a wide value range with high measuring accuracy both value pairs should be used for calibra...
Page 28: ...imension and note it down In the case of fluctuating values carry out several measurements and average the individual raw values 7 Take a sample at the same time as the measurement and as close to the...
Page 29: ...nd value pair higher concentration or raw value of the single point calibration Cal ref value 2 0 00 20000 00 Reference value of the second value pair higher concentration or reference value of the si...
Page 30: ...location of the Zero adjustment data The data of the sensor checks are not stored The data of the user zero adjustment are stored in the sensor Outlet A successful Sensor check H2O dest is the prereq...
Page 31: ...ating error or defective sensor Sensor blocked for measurement Sensor check Standard OK Error 40 40 40 or 40 Operating error or defective sensor Zero adjustment Error OK After several zero adjustments...
Page 32: ...g invalid measured value 4 Confirm the maintenance condition with OK 5 Take the sensor out of the test sample 6 Remove the blind plug or unscrew the compressed air connection from the sensor 7 Thoroug...
Page 33: ...ll the measuring chamber with standard solution see section 6 12 If there are any air bubbles on the measuring windows remove them by gently knocking on the sensor sleeve with a bar Carry out the inte...
Page 34: ...tion data The maintenance condition however has to be switched off again in any case After starting the calibration with the OK key step 16 you can no longer cancel the calibration procedure 16 Confir...
Page 35: ...ul Error Sensor check standard not successful Possible cause wrong or expired standard 17 Confirm the calibration result with OK 18 Confirm the message After successful cal with OK The display returns...
Page 36: ...e sensor with ultrapure water Accessories Sponge Plastic bristled brushes of different sizes Interdental brushes are suitable for the cleaning of the compressed air chan nels Flocked cleaning cards se...
Page 37: ...djustment should be carried out must run through all cleaning steps properly During the cleaning process always keep all surfaces moist to prevent already dissolved contamination and detergents from d...
Page 38: ...able and mounting parts Note Follow the instruction manual of the disinfectant reaction time 4 Remove the blind plug or unscrew the compressed air connection from the sensor 5 Clean the sensor with so...
Page 39: ...ry leave the cleaning card in the measuring gap for a few minutes so the detergent can take effect Thoroughly rinse with ultrapure water the measuring gap and the surrounding areas moistened with dete...
Page 40: ...low concentration for 5 mm UVVIS SAC NOx sensors ST 1 481 077 Standard 2 with high concentration for 1 mm UVVIS SAC NOx sensors ST 2 481 078 Description Model Order no Compressed air cleaning set with...
Page 41: ...o Mounting set for horizontal installation with EH F 170 swing mounting assembly VIS Set EH 481 073 Set for the mounting of UVVIS SAC NOx sensors to the S 200 electrode floater VIS Set F 481 080 Flow...
Page 42: ...4 3 3 Zero adjustment still erroneous after several trials Contact service Unknown See log book Cause Remedy Erroneous user calibration Repeat user calibration Make sure the samples for comparison are...
Page 43: ...surement windows section 5 2 4 Check all components of the com pressed air cleaning system com pressor compressed air line sensor rinsing openings Substantial change of test sample sample matrix Carry...
Page 44: ...x IQ SAC too long voltage drop too great Install another power supply module in the vicinity of the sensor Electrical connection between power supply module and UV 70x IQ SAC is interrupted Check the...
Page 45: ...AC 5 mm Measuring mode Measuring range Resolution SAC 254 dissolved 0 0 3000 1 m 1 1 m SAC 254 total 0 0 3000 1 m 1 1 m UVT 254 total The UVT 254 value is standardized to 10 mm gap width 0 0 100 0 0 1...
Page 46: ...tandardized to 10 mm gap width The quoted measuring ranges are nominal measuring ranges that are theoretically possible In practice real measuring ranges exist that are given by the limits of photomet...
Page 47: ...891 hock protection 60 96 891 120 Socket SACIQ With shock protection 774 865 381 Approx 3 8 kg without shock protectors and without sensor connec tion cable Approx 4 8 kg with shock protectors but wi...
Page 48: ...L 61010 1 Certified to CAN CSA C22 2 61010 1 Nominal voltage Max 24VDC via the IQ SENSOR NET for more details see chapter TECHNICAL DATA of the IQ SENSOR NET system operating manual Power consumption...
Page 49: ...or Some error messages contain an internal error code starting with Information on the contents and structure of the log book and the structure of the message code is given in the LOG BOOK chapter of...
Page 50: ...apure water Repeat the zero adjustment EAM3Ax Meas range exceeded or undercut Check process EC33Ax User calibration erroneous Check raw value reference value pairs Set all values within the measuring...
Page 51: ...ow optical measurement Execute zero ajustment ES5541 Failure optical measurement hardware Contact service ES6541 Failure optical measurement software Contact service ES7541 Failure optical measurement...
Page 52: ...ake sure the sensor operates correctly IC4541 The last valid user zero adjustment was activated Make sure the sen sor operates correctly IC5541 Invalid user zero adjustment was replaced by last valid...
Page 53: ...ings Service List of all compo nents menu see system operating manual by an automated query from a superordinate process control e g when connected to the Profi bus of the IQ Data Server see operating...
Page 54: ...Mass or amount of a dissolved substance per volume e g in g L or mol L CSB COD Abbreviation for the Chemical Oxygen Demand The chemical oxy gen demand is the equivalent amount of oxygen that is requir...
Page 55: ...ty correction reference measurement at 550 nm SAK SAC total Spectral absorption coefficient at a certain wavelength The SAC 254 value is the photometrically measured absorbance of a test sample at a w...
Page 56: ...Appendix Glossary UV 70x IQ SAC 56 ba77046e02 05 2017...
Page 57: ......
Page 58: ...rove the way water is used conserved and re used in the future is central to our work Our products and services move treat analyze monitor and return water to the environment in public utility industr...