pHotoFlex
®
Turb
Operation
35
ba75976e07
09/2021
Display when the
measuring range
is exceeded or
undercut
Examples:
5 Start the measurement with
<START/ENTER>
.
Measurement is started. The result is displayed.
Photometry
\
Concentration
[
BV
]
0.74
mg/l
83: A6/25 MC
NH4-N
16 mm
0.20 - 6.51 mg/l
01.02.05 15:12
A blank value measured by
the user is used
Display
Meaning
"< [Lower limit of measuring
range]"
instead of the measured value
Measuring range undercut.
Remedy: Select a test with a lower
measuring range
"> [Upper limit of measuring
range]"
instead of the measured value
Measuring range exceeded.
Remedy: Select a test with a higher
measuring range or dilute the sample
Orientation value
This value serves as an indication for
the selection of a suitable test or dilu-
tion.
If an evaluation of the measurement
signal is no longer possible (intensity
too high or too low), four bars appear
("----").
Photometry
\
Concentration
0.16
[
BV
]
< 0.20
mg/l
83: A6/25 MC
NH4-N
16 mm
0.20 - 6.51 mg/l
01.02.05 15:12
Orientation value