D.O. calibration
Why calibrate?
D.O. sensors age. This changes the slope of the D.O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
Calibration
/
Calibration interval
1 ... 999 d
Calibration interval
for the
D.O. sensor (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring screen.
Calibration
/
Comparison meas.
On
Off
Enables to adjust the mea-
sured value with the aid of a
comparison measurement,
e.g. Winkler titration.
For details, see section
4.4.4.
Sal correction
On
Off
Manual salt content correc-
tion for concentration mea-
surements.
Salinity
0.0 ... 70.0
Salinity or salinity equivalent
for the salt content correc-
tion.
This function is only avail-
able for concentration mea-
surements if the manual salt
content correction is
switched on.
Stability control
On
Off
Switches on or off the auto-
matic stability control during
measurement (see section
4.3.3)
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperature values are
displayed with the selected
unit.
Reset
-
Resets all sensor settings to
the delivery condition (see
section 4.8.1).
Menu item
Possible
setting
Explanation
Summary of Contents for WTW Oxi 3310
Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Page 63: ......